SwePub
Sök i LIBRIS databas

  Utökad sökning

id:"swepub:oai:DiVA.org:uu-21801"
 

Sökning: id:"swepub:oai:DiVA.org:uu-21801" > Variations in assay...

Variations in assay protocol for the Dako cystatin C method may change patient results by 50% without changing the results for controls

Flodin, Mats (författare)
Uppsala universitet,Klinisk kemi,Biokemisk struktur och funktion
Hansson, Lars-Olof (författare)
Uppsala universitet,Klinisk kemi,Biokemisk struktur och funktion
Larsson, Anders (författare)
Uppsala universitet,Klinisk kemi,Biokemisk struktur och funktion
 (creator_code:org_t)
2006
2006
Engelska.
Ingår i: Clinical Chemistry and Laboratory Medicine. - 1434-6621 .- 1437-4331. ; 44:12, s. 1481-1485
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Background: Cystatin C is increasingly used as a glomerular filtration marker, but so far only a few companies produce most of the cystatin C reagents suited for turbidimetry or nephelometry use in clinical laboratories. Methods: We studied different protocols for measuring cystatin C on an Architect ci8200 system using cystatin C reagents from Dako (Glostrup, Denmark). The results were compared with those obtained with Dade Behring reagents (Deerfield, IL, USA) on a BN ProSpec system. Results: Differences in assay protocol on the same instrument with the Dako reagent yielded an up to 50% difference in glomerular filtration rate calculated from the cystatin C results when analyzing patient samples, but had no effect on the results for controls. There were also significant differences regarding linearity and kinetics between samples and controls/calibrators. Conclusions: The results indicate different reactivity of the Dako antibodies against calibrators and controls in comparison with patient samples, highlighting the importance of using controls and calibrators that do not differ from patient samples.

Nyckelord

Cystatin C
Erroneous results
Glomerular filtration ratio
Immunoassay
Kidney function
Recombinant proteins
Turbidimetry
MEDICINE
MEDICIN

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Hitta mer i SwePub

Av författaren/redakt...
Flodin, Mats
Hansson, Lars-Ol ...
Larsson, Anders
Artiklar i publikationen
Clinical Chemist ...
Av lärosätet
Uppsala universitet

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy