Automatic extraction of forward stroke volume using dynamic PET/CT : a dual-tracer and dual-scanner validation in patients with heart valve disease.

Harms, Hendrik Johannes (författare)
Department of Nuclear Medicine & PET Centre, Aarhus University Hospital
Tolbod, Lars Poulsen (författare)
Department of Nuclear Medicine & PET Centre, Aarhus University Hospital
Hansson, Nils Henrik Stubkjær (författare)
Aarhus Univ Hosp, Dept Cardiol, DK-8200 Aarhus N, Denmark
visa fler...
Kero, Tanja (författare)
Uppsala universitet,Radiologi
Örndahl, Lovisa Holm (författare)
Uppsala universitet,Kardiologi
Kim, Won Yong (författare)
Aarhus Univ Hosp, Dept Cardiol, DK-8200 Aarhus N, Denmark
Bjerner, Tomas (författare)
Uppsala universitet,Radiologi
Bouchelouche, Kirsten (författare)
Department of Nuclear Medicine & PET Centre, Aarhus University Hospital
Wiggers, Henrik (författare)
Aarhus Univ Hosp, Dept Cardiol, DK-8200 Aarhus N, Denmark
Frøkiær, Jørgen (författare)
Department of Nuclear Medicine & PET Centre, Aarhus University Hospital
Sörensen, Jens (författare)
Uppsala universitet,Radiologi,Department of Nuclear Medicine & PET Centre, Aarhus University Hospital
visa färre...
 (creator_code:org_t)
2015-10-26
2015
Engelska.
Ingår i: EJNMMI Physics. - : Springer Science and Business Media LLC. - 2197-7364. ; 2:1
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • BACKGROUND: The aim of this study was to develop and validate an automated method for extracting forward stroke volume (FSV) using indicator dilution theory directly from dynamic positron emission tomography (PET) studies for two different tracers and scanners.METHODS: 35 subjects underwent a dynamic (11)C-acetate PET scan on a Siemens Biograph TruePoint-64 PET/CT (scanner I). In addition, 10 subjects underwent both dynamic (15)O-water PET and (11)C-acetate PET scans on a GE Discovery-ST PET/CT (scanner II). The left ventricular (LV)-aortic time-activity curve (TAC) was extracted automatically from PET data using cluster analysis. The first-pass peak was isolated by automatic extrapolation of the downslope of the TAC. FSV was calculated as the injected dose divided by the product of heart rate and the area under the curve of the first-pass peak. Gold standard FSV was measured using phase-contrast cardiovascular magnetic resonance (CMR).RESULTS: FSVPET correlated highly with FSVCMR (r = 0.87, slope = 0.90 for scanner I, r = 0.87, slope = 1.65, and r = 0.85, slope = 1.69 for scanner II for (15)O-water and (11)C-acetate, respectively) although a systematic bias was observed for both scanners (p < 0.001 for all). FSV based on (11)C-acetate and (15)O-water correlated highly (r = 0.99, slope = 1.03) with no significant difference between FSV estimates (p = 0.14).CONCLUSIONS: FSV can be obtained automatically using dynamic PET/CT and cluster analysis. Results are almost identical for (11)C-acetate and (15)O-water. A scanner-dependent bias was observed, and a scanner calibration factor is required for multi-scanner studies. Generalization of the method to other tracers and scanners requires further validation.

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy