The Roche Immunoturbidimetric Albumin Method on Cobas c 501 Gives Higher Values Than the Abbott and Roche BCP Methods When Analyzing Patient Plasma Samples

Helmersson-Karlqvist, Johanna (författare)
Uppsala universitet,Biokemisk struktur och funktion
Flodin, Mats (författare)
Uppsala universitet,Biokemisk struktur och funktion
Havelka, Aleksandra Mandic (författare)
Karolinska Univ Hosp, Dept Clin Chem, Stockholm, Sweden
visa fler...
Xu, Xiao Yan (författare)
Uppsala universitet,Biokemisk struktur och funktion
Larsson, Anders (författare)
Uppsala universitet,Biokemisk struktur och funktion
visa färre...
 (creator_code:org_t)
2016-05-12
2016
Engelska.
Ingår i: Journal of clinical laboratory analysis (Print). - : Wiley. - 0887-8013 .- 1098-2825. ; 30:5, s. 677-681
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • BACKGROUND: Serum/plasma albumin is an important and widely used laboratory marker and it is important that we measure albumin correctly without bias. We had indications that the immunoturbidimetric method on Cobas c 501 and the bromocresol purple (BCP) method on Architect 16000 differed, so we decided to study these methods more closely.METHOD: A total of 1,951 patient requests with albumin measured with both the Architect BCP and Cobas immunoturbidimetric methods were extracted from the laboratory system. A comparison with fresh plasma samples was also performed that included immunoturbidimetric and BCP methods on Cobas c 501 and analysis of the international protein calibrator ERM-DA470k/IFCC.RESULTS: The median difference between the Abbott BCP and Roche immunoturbidimetric methods was 3.3 g/l and the Roche method overestimated ERM-DA470k/IFCC by 2.2 g/l. The Roche immunoturbidimetric method gave higher values than the Roche BCP method: y = 1.111x - 0.739, R² = 0.971.CONCLUSION: The Roche immunoturbidimetric albumin method gives clearly higher values than the Abbott and Roche BCP methods when analyzing fresh patient samples. The differences between the two methods were similar at normal and low albumin levels.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Klinisk laboratoriemedicin (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Clinical Laboratory Medicine (hsv//eng)

Nyckelord

albumin; BCP; bromocresol purple; calibration; immunoturbidimetric method; turbidimetry

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy