Magnetic Resonance Compatibility of a Transponder Aimed for Radiotherapy Positioning – A Phantom Study

Wikström, Johan, 1964- (författare)
Uppsala universitet,Radiologi,Department of Surgical Sciences, Section of Radiology, Uppsala University, Uppsala, Sweden
Isacsson, Ulf (författare)
Uppsala universitet,Medicinsk strålningsvetenskap,Department of Immunology, Genetics and Pathology, Section of Medical Radiation Sci-ence, Uppsala University, Uppsala, Sweden
Johansson, Bengt, 1958- (författare)
Örebro universitet,Institutionen för medicinska vetenskaper,Region Örebro län,Department of Oncology, Örebro University Hospital, Örebro, Sweden
visa fler...
Lennernäs, Bo, 1963- (författare)
Örebro universitet,Institutionen för medicinska vetenskaper,Region Örebro län,Department of Oncology, Örebro University Hospital, Örebro, Sweden
visa färre...
 (creator_code:org_t)
Anticancer Research USA Inc. 2017
2017
Engelska.
Ingår i: Anticancer Research. - : Anticancer Research USA Inc.. - 0250-7005 .- 1791-7530. ; 37:9, s. 4993-4996
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Background/Aim: Electromagnetic Positioning Systems (EMP) is a new position-ing technique in four-dimensional radiotherapy. Patients with implanted transponders may be referred for magnetic resonance imaging (MRI) making it important to establish the MR safety.Materials and Methods: Oranges were prepared with transponders and imaged on a 3T MR scanner with different sequences. Computed tomography (CT) was performed as comparison. MR artifacts were assessed. An estimation of the maximum transponder de-flection force and heating was made.Results: The mean measured displacement of transponders was 0.1 mm (range=0.03-0.3 mm). Artifacts were observed adjacent to transponders using all sequences. The deflection force on the transponder in the gantry was less than 38 mN. No heating was observed.Conclusion: The absence of any substantial movement, the weak measured deflection force and absence of observed heating speaks for the safe use of MR imaging with transponder 3T. Local artefacts makes evaluation impossible adjacent to transponders.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Cancer och onkologi (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Cancer and Oncology (hsv//eng)

Nyckelord

Prostate cancer
electromagnetic positioning
four dimensional radiotherapy

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy