SwePub
Sök i LIBRIS databas

  Utökad sökning

id:"swepub:oai:DiVA.org:uu-404495"
 

Sökning: id:"swepub:oai:DiVA.org:uu-404495" > Low-dose CT-based i...

Low-dose CT-based implant motion analysis is a precise tool for early migration measurements of hip cups : a clinical study of 24 patients

Brodén, Cyrus (författare)
Imperial Coll London, Dept Surg & Canc, London, England; Karolinska Inst, Danderyd Hosp, Dept Clin Sci, Stockholm, Sweden
Sandberg, Olof (författare)
‎ Sectra, Linköping, Sweden
Sköldenberg, Olof (författare)
Karolinska Institutet
visa fler...
Stigbrand, Hampus (författare)
Uppsala universitet,Centrum för klinisk forskning, Gävleborg,Länssjukhuset, Dept Orthoped Surg, Gävle, Sweden
Hänni, Mari (författare)
Uppsala universitet,Radiologi
Giles, Joshua W (författare)
Univ Victoria, Dept Mech Engn, Victoria, BC, Canada
Emery, Roger (författare)
St Marys Hosp, Dept Orthopaed Surg, London, England
Lazarinis, Stergios, 1971- (författare)
Uppsala universitet,Ortopedi
Nyström, Andreas (författare)
Uppsala universitet,Ortopedi
Olivecrona, Henrik (författare)
Karolinska Inst, Dept Mol Med & Surg, Stockholm, Sweden
visa färre...
 (creator_code:org_t)
2020-02-14
2020
Engelska.
Ingår i: Acta Orthopaedica. - : Informa UK Limited. - 1745-3674 .- 1745-3682. ; 91:3, s. 260-265
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Background and purpose - Early implant migration is known to be a predictive factor of clinical loosening in total hip arthroplasty (THA). Radiostereometric analysis (RSA) is the gold standard used to measure early migration in patients. However, RSA requires costly, specialized imaging equipment and the image process is complex. We determined the precision of an alternative, commercially available, CT method in 3 ongoing clinical THA studies, comprising 3 different cups.Materials and methods - 24 CT double examinations of 24 hip cups were selected consecutively from 3 ongoing prospective studies: 2 primary THA (1 cemented and 1 uncemented) and 1 THA (cemented) revision study. Precision of the CT-based implant motion analysis (CTMA) system was calculated separately for each study, using both the surface anatomy of the pelvis and metal beads placed in the pelvis.Results - For the CTMA analysis using the surface anatomy of the pelvis, the precision ranged between 0.07 and 0.31 mm in translation and 0.20° and 0.39° for rotation, respectively. For the CTMA analysis using beads the precision ranged between 0.08 and 0.20 mm in translation and between 0.20° and 0.43° for rotations. The radiation dose ranged between 0.2 and 2.3 mSv.Interpretation - CTMA achieved a clinically relevant and consistent precision between the 3 different hip cups studied. The use of different hip cup types, different CT scanners, or registration method (beads or surface anatomy) had no discernible effect on precision. Therefore, CTMA without the use of bone markers could potentially be an alternative to RSA to measure early migration.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Ortopedi (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Orthopaedics (hsv//eng)

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy