Evaluation of Arterial Spin Labeling MRI : Comparison with 15O-Water PET on an Integrated PET/MR Scanner

Fahlström, Markus (författare)
Uppsala universitet,Radiologi
Appel, Lieuwe (författare)
Uppsala universitet,Radiologi
Kumlien, Eva (författare)
Uppsala universitet,Landtblom: Neurovetenskap
visa fler...
Danfors, Torsten, 1964- (författare)
Uppsala universitet,Radiologi
Engström, Mathias (författare)
GE Healthcare, Applied Science Laboratory
Wikström, Johan, 1964- (författare)
Uppsala universitet,Radiologi
Antoni, Gunnar (författare)
Uppsala universitet,Preparativ läkemedelskemi
Larsson, Elna-Marie (författare)
Uppsala universitet,Radiologi
Lubberink, Mark (författare)
Uppsala universitet,Radiologi
visa färre...
 (creator_code:org_t)
2021-05-01
2021
Engelska.
Ingår i: Diagnostics (Basel). - : MDPI. - 2075-4418. ; 11:5
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Cerebral blood flow (CBF) measurements are of high clinical value and can be acquired non-invasively with no radiation exposure using pseudo-continuous arterial spin labeling (ASL). The aim of this study was to evaluate accordance in resting state CBF between ASL (CBFASL) and 15O-water positron emission tomography (PET) (CBFPET) acquired simultaneously on an integrated 3T PET/MR system. The data comprised ASL and dynamic 15O-water PET data with arterial blood sampling of eighteen subjects (eight patients with focal epilepsy and ten healthy controls, age 21 to 61 years). 15O-water PET parametric CBF images were generated using a basis function implementation of the single tissue compartment model. Cortical and subcortical regions were automatically segmented using Freesurfer. Average CBFASL and CBFPET in grey matter were 60 ± 20 and 75 ± 22 mL/100 g/min respectively, with a relatively high correlation (r = 0.78, p < 0.001). Bland-Altman analysis revealed poor agreement (bias = −15 mL/100 g/min, lower and upper limits of agreements = −16 and 45 mL/100 g/min, respectively) with a negative relationship. Accounting for the negative relationship, the width of the limits of agreement could be narrowed from 61 mL/100 g/min to 35 mL/100 g/min using regression-based limits of agreements. Although a high correlation between CBFASL and CBFPET was found, the agreement in absolute CBF values was not sufficient for ASL to be used interchangeably with 15O-water PET.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Nyckelord

ASL
CBF
PET/MR
validation
15O-water

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy