SwePub
Sök i LIBRIS databas

  Extended search

id:"swepub:oai:DiVA.org:uu-407607"
 

Search: id:"swepub:oai:DiVA.org:uu-407607" > Validating the insp...

  • 1 of 1
  • Previous record
  • Next record
  •    To hitlist

Validating the inspired sinewave technique to measure the volume of the 'baby lung' in a porcine lung-injury model

Crockett, Douglas C. (author)
Univ Oxford, Nuffield Div Anaesthet, Oxford, England
Tran, Minh C. (author)
Univ Oxford, Nuffield Div Anaesthet, Oxford, England;Univ Oxford, Dept Engn Sci, Oxford, England
Formenti, Federico (author)
Univ Oxford, Nuffield Div Anaesthet, Oxford, England;Kings Coll London, Ctr Human & Appl Physiol Sci, London, England;Univ Nebraska, Dept Biomech, Omaha, NE 68182 USA
show more...
Cronin, John N. (author)
Kings Coll London, Ctr Human & Appl Physiol Sci, London, England
Hedenstierna, Göran, 1941- (author)
Uppsala universitet,Klinisk fysiologi,Hedenstiernalaboratoriet
Larsson, Anders (author)
Uppsala universitet,Anestesiologi och intensivvård
Phan, Phi A. (author)
Univ Oxford, Nuffield Div Anaesthet, Oxford, England
Farmery, Andrew D. (author)
Univ Oxford, Nuffield Div Anaesthet, Oxford, England
show less...
 (creator_code:org_t)
ELSEVIER SCI LTD, 2020
2020
English.
In: British Journal of Anaesthesia. - : ELSEVIER SCI LTD. - 0007-0912 .- 1471-6771. ; 124:3, s. 345-353
  • Journal article (peer-reviewed)
Abstract Subject headings
Close  
  • Background: Bedside lung volume measurement could personalise ventilation and reduce driving pressure in patients with acute respiratory distress syndrome (ARDS). We investigated a modified gas-dilution method, the inspired sinewave technique (IST), to measure the effective lung volume (ELV) in pigs with uninjured lungs and in an ARDS model. Methods: Anaesthetised mechanically ventilated pigs were studied before and after surfactant depletion by saline lavage. Changes in PEEP were used to change ELV. Paired measurements of absolute ELV were taken with IST (ELVIST) and compared with gold-standard measures (sulphur hexafluoride wash in/washout [ELVSF6] and computed tomography (CT) [ELVCT]). Measured volumes were used to calculate changes in ELV (Delta ELV) between PEEP levels for each method (Delta ELVIST, Delta ELVSF6, and Delta ELVCT). Results: The coefficient of variation was <5% for repeated ELVIST measurements (n=13 pigs). There was a strong linear relationship between ELVIST and ELVSF6 in uninjured lungs (r(2)=0.97), and with both ELVSF6 and ELVCT in the ARDS model (r(2)=0.87 and 0.92, respectively). ELVIST had a mean bias of -12 to 13% (95% limits=+/- 17 - 25%) compared with ELVSF6 and ELVCT. Delta ELVIST was concordant with Delta ELVSF6 and Delta ELVCT in 98-100% of measurements, and had a mean bias of -73 to -77 ml (95% limits=+/- 128 - 186 ml) compared with Delta ELVSF6 and -1 ml (95% limits +/- 333 ml) compared with Delta ELVCT. Conclusions: IST provides a repeatable measure of absolute ELV and shows minimal bias when tracking PEEP-induced changes in lung volume compared with CT in a saline-lavage model of ARDS.

Subject headings

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Anestesi och intensivvård (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Anesthesiology and Intensive Care (hsv//eng)

Keyword

ARDS
computed tomography
inspired sinewave technique
lung volume measurement
method comparison

Publication and Content Type

ref (subject category)
art (subject category)

Find in a library

To the university's database

  • 1 of 1
  • Previous record
  • Next record
  •    To hitlist

Search outside SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Close

Copy and save the link in order to return to this view