Safety and efficacy evaluation of the automatic stepwise recruitment maneuver in the neonatal population : An in vivo interventional study. Can anesthesiologists safely perform automatic lung recruitment maneuvers in neonates?

Serrano Zueras, Clara (författare)
Hosp Univ Puerta Hierro Majadahonda, Dept Anaesthesiol Intens Care & Pain, Majadahonda, Spain.
Guilló Moreno, Verónica (författare)
Hosp Univ Puerta Hierro Majadahonda, Dept Anaesthesiol Intens Care & Pain, Majadahonda, Spain.
Santos González, Martín (författare)
Hosp Univ Puerta Hierro Majadahonda, Med & Surg Res Unit, Inst Invest Sanitaria Puerta Hierro Segovia Arana, Majadahonda, Spain.
visa fler...
Gómez Nieto, Francisco Javier (författare)
Hosp Univ Puerta Hierro Majadahonda, Dept Anaesthesiol Intens Care & Pain, Majadahonda, Spain.
Hedenstierna, Göran, 1941- (författare)
Uppsala universitet,Klinisk fysiologi
García Fernández, Javier (författare)
Hosp Univ Puerta Hierro Majadahonda, Dept Anaesthesiol Intens Care & Pain, Majadahonda, Spain.
visa färre...
Hosp Univ Puerta Hierro Majadahonda, Dept Anaesthesiol Intens Care & Pain, Majadahonda, Spain Hosp Univ Puerta Hierro Majadahonda, Med & Surg Res Unit, Inst Invest Sanitaria Puerta Hierro Segovia Arana, Majadahonda, Spain. (creator_code:org_t)
2021-07-06
2021
Engelska.
Ingår i: Pediatric Anaesthesia. - : John Wiley & Sons. - 1155-5645 .- 1460-9592. ; 31:9, s. 1003-1010
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • BackgroundA new software has recently been incorporated in almost all new anesthesia machines to enable automatic lung recruitment maneuvers. To date, no studies have assessed the safety and efficacy of these automatic software programs in the neonatal population.AimsWe aimed to evaluate the safety and efficacy of the lung recruitment maneuver performed using the automatic stepwise recruitment maneuver software of the FLOW-i 4.3 Anesthesia System® in a healthy and live neonatal model.MethodsEight male newborn piglets were included in the study. The lung recruitment maneuver was performed in pressure-controlled ventilation with a constant driving pressure (15 cmH2O) in a stepwise increasing positive end-expiratory pressure (PEEP) model. The target peak inspiratory pressure (PIP) was 30 cmH2O and PEEP was 15 cmH2O. The maneuver lasted for 39 seconds. The hemodynamic variables were monitored using the PICCO® system. The following respiratory parameters were monitored: oxygen saturation, fraction of inspired oxygen, partial pressure of oxygen and carbon dioxide in the arterial blood, end-tidal carbon dioxide pressure, PIP, plateau pressure, PEEP, static compliance (Cstat), and dynamic compliance (Cdyn).Safety was evaluated by assessing the accuracy of the software, need for not interrupting the maneuver, hemodynamic stability, and absence of adverse respiratory events with the lung recruitment maneuver. Efficacy was evaluated by improvement in Cstat and Cdyn after performing the lung recruitment maneuver.ResultsAll lung recruitment maneuvers were safely performed as scheduled without any interruptions. No pneumothorax or other side effects were observed. Hemodynamic stability was maintained during the lung recruitment maneuver. We observed an improvement of 33% in Cdyn and 24% in Cstat after the maneuver.ConclusionsThe automatic stepwise recruitment maneuver software of the FLOW-i 4.3 Anesthesia System® is safe and efficacious in a healthy neonatal model. We did not observe any adverse respiratory or hemodynamic events during the implementation of the lung recruitment maneuver in the pressure-controlled ventilation mode using a stepwise increasing PEEP (30/15 cmH2O) approach.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Anestesi och intensivvård (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Anesthesiology and Intensive Care (hsv//eng)

Nyckelord

automatic stepwise recruitment maneuver
mechanical ventilation
neonates
recruitment maneuver
safety

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy