Optimisation of tube voltage for conventional urography using a Gd2O2S:Tb flat panel detector

Zachrisson, Sara (author)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiofysik,Institute of Clinical Sciences, Department of Radiation Physics
Hansson, Jonny (author)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiofysik,Institute of Clinical Sciences, Department of Radiation Physics
Cederblad, Åke, 1947 (author)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiofysik,Institute of Clinical Sciences, Department of Radiation Physics
show more...
Geterud, Kjell (author)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiologi,Institute of Clinical Sciences, Department of Radiology
Båth, Magnus, 1974 (author)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiofysik,Institute of Clinical Sciences, Department of Radiation Physics
show less...
 (creator_code:org_t)
2010-03-11
2010
English.
In: Radiation protection dosimetry. - : Oxford University Press (OUP). - 1742-3406 .- 0144-8420. ; 139:1-3, s. 86-91
  • Journal article (peer-reviewed)
Abstract Subject headings
Close  
  • With the increasing use of computed tomography (CT) for urography examinations, the indications for 'conventional' projection urography have changed and are more focused on high-contrast details. The purpose of the present study was to optimise the beam quality for urography examinations performed with a Gd(2)O(2)S:Tb flat-panel detector for the new conditions. Images of an anthropomorphic phantom were collected at different tube voltages with a CXDI-40G detector (Canon Inc., Tokyo, Japan). The images were analysed by radiologists and residents in a visual grading characteristics (VGCs) study. The tube voltage resulting in the best image quality was 55 kV, which therefore was selected for a clinical study. Images from 62 patients exposed with either 55 or 73 kV (original tube voltage) at constant effective doses were included. The 55-kV images underwent simulated dose reduction to represent images collected at 80, 64, 50, 40 and 32 % of the original dose level. All images were included in a VGC study where the observers rated the visibility of important anatomical landmarks. For images collected at 55 kV, an effective dose of approximately 85 % resulted in the same image quality as for images collected at 73 kV at 100 % dose. In conclusion, a low tube voltage should be used for conventional urography focused on high-contrast details. The study indicates that using a tube voltage of 55 kV instead of 73 kV for a Gd(2)O(2)S:Tb flat-panel detector, the effective dose can be reduced by approximately 10-20 % for normal-sized patients while maintaining image quality.

Subject headings

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Publication and Content Type

ref (subject category)
art (subject category)

Find in a library

To the university's database

  • 1 of 1
  • Previous record
  • Next record
  •    To hitlist

Search outside SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Close

Copy and save the link in order to return to this view