SwePub
Sök i LIBRIS databas

  Utökad sökning

id:"swepub:oai:gup.ub.gu.se/124093"
 

Sökning: id:"swepub:oai:gup.ub.gu.se/124093" > Contrast agent infl...

Contrast agent influences MRI phase-contrast flow measurements in small vessels.

Lagerstrand, Kerstin M (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiofysik,Institute of Clinical Sciences, Department of Radiation Physics
Vikhoff-Baaz, Barbro (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiofysik,Institute of Clinical Sciences, Department of Radiation Physics
Starck, Göran (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiofysik,Institute of Clinical Sciences, Department of Radiation Physics
visa fler...
Forssell-Aronsson, Eva, 1961 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiofysik,Institute of Clinical Sciences, Department of Radiation Physics
visa färre...
 (creator_code:org_t)
2010-05-14
2010
Engelska.
Ingår i: Magnetic resonance in medicine. - : Wiley. - 1522-2594 .- 0740-3194. ; 64:1, s. 42-46
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Contrast-enhanced MR angiography is often combined with phase contrast (PC) flow measurement to answer a particular clinical question. The contrast agent that is administered during contrast-enhanced MR angiography may still be present in the blood during the consecutive PC flow measurement. The aim of this work was to evaluate the influence of contrast agent on PC flow measurements in small vessels. For that purpose, both in vivo measurements and computer simulations were performed. The dependence of the PC flow quantification on the signal amplitude difference between blood and stationary background tissue for various vessel sizes was characterized. Results show that the partial-volume effect strongly affects the accuracy of the PC flow quantification when the imaged vessel is small compared to the spatial resolution. A higher blood-to-background-contrast level during imaging significantly increases the partial-volume effect and thereby reduces the accuracy of the flow quantification. On the other hand, a higher blood-to-background-contrast level facilitated the segmentation of the vessel for flow rate determination. PC flow measurements should therefore be performed after contrast agent administration in large vessels, but before contrast agent administration in small vessels.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Nyckelord

magnetic resonance imaging
phase-contrast
spatial resolution
contrast agent
spatial resolution

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy