Dose optimisation of double-contrast barium enema examinations.

Berner, Karin (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiofysik,Institute of Clinical Sciences, Department of Radiation Physics
Båth, Magnus, 1974 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiofysik,Institute of Clinical Sciences, Department of Radiation Physics
Jonasson, Pernilla (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiofysik,Institute of Clinical Sciences, Department of Radiation Physics
visa fler...
Cappelen-Smith, J (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiologi,Institute of Clinical Sciences, Department of Radiology
Fogelstam, P (författare)
Söderberg, Jonas (författare)
visa färre...
 (creator_code:org_t)
2010-03-15
2010
Engelska.
Ingår i: Radiation protection dosimetry. - : Oxford University Press (OUP). - 1742-3406 .- 0144-8420. ; 139:1-3, s. 388-392
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • The purpose of the present work was to optimise the filtration and dose setting for double-contrast barium enema examinations using a Philips MultiDiagnost Eleva FD system. A phantom study was performed prior to a patient study. A CDRAD phantom was used in a study where copper and aluminium filtration, different detector doses and tube potentials were examined. The image quality was evaluated using the software CDRAD Analyser and the phantom dose was determined using the Monte Carlo-based software PCXMC. The original setting [100 % detector dose (660 nGy air kerma) and a total filtration of 3.5 mm Al, at 81 kVp] and two other settings identified by the phantom study (100 % detector dose and additional filtration of 1 mm Al and 0.2 mm Cu as well as 80 % detector dose and added filtration of 1 mm Al and 0.2 mm Cu) were included in the patient study. The patient study included 60 patients and up to 8 images from each patient. Six radiologists performed a visual grading characteristics study to evaluate the image quality. A four-step scale was used to judge the fulfillment of three image quality criteria. No overall statistical significant difference in image quality was found between the three settings (P > 0.05). The decrease in the effective dose for the settings in the patient study was 15 % when filtration was added and 34 % when both filtrations was added and detector dose was reduced. The study indicates that additional filtration of 1 mm Al and 0.2 mm Cu and a decrease in detector dose by 20 % from the original setting can be used in colon examinations with Philips MultiDiagnost Eleva FD to reduce the patient dose by 30 % without significantly affecting the image quality. For 20 exposures, this corresponds to a decrease in the effective dose from 1.6 to 1.1 mSv.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy