Feedback Analysis in Percutaneous Bone-Conduction Device and Bone-Conduction Implant on a Dry Cranium
-
- Taghavi, Hamidreza, 1981 (författare)
- Chalmers tekniska högskola,Chalmers University of Technology
-
- Håkansson, Bo, 1953 (författare)
- Chalmers tekniska högskola,Chalmers University of Technology
-
- Reinfeldt, Sabine, 1978 (författare)
- Chalmers tekniska högskola,Chalmers University of Technology
-
visa fler...
-
- Eeg-Olofsson, Måns, 1967 (författare)
- Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för öron-, näs- och halssjukdomar,Institute of Clinical Sciences, Department of Otorhinolaryngology,University of Gothenburg
-
- Akhshijan, Shirin Sadat, 1985 (författare)
- Chalmers tekniska högskola,Chalmers University of Technology
-
visa färre...
-
(creator_code:org_t)
- 2012
- 2012
- Engelska.
-
Ingår i: Otology and Neurotology. - 1531-7129 .- 1537-4505. ; 33:3, s. 413-420
- Relaterad länk:
-
https://gup.ub.gu.se... (primary) (free)
-
visa fler...
-
http://publications.... (primary) (free)
-
https://gup.ub.gu.se...
-
https://doi.org/10.1...
-
https://research.cha...
-
visa färre...
Abstract
Ämnesord
Stäng
- Hypothesis: The bone-conduction implant (BCI) can use a higher gain setting without having feedback problems compared with a percutaneous bone-conduction device (PBCD). Background: The conventional PBCD, today, is a common treatment for patients with conductive hearing loss and single-sided deafness. However, there are minor drawbacks reported related to the percutaneous implant and specifically poor high-frequency gain. The BCI system is designed as an alternative to the percutaneous system because it leaves the skin intact and is less prone to fall into feedback oscillations, thus allowing more high-frequency gain. Methods: Loop gains of the Baha Classic 300 and the BCI were measured in the frequency range of 100 to 10,000 Hz attached to a Skull simulator and a dry cranium. The Baha and the BCI positions were investigated. The devices were adjusted to full-on gain. Results: It was found that the gain headroom using the BCI was generally 0 to 10 dB better at higher frequencies than using the Baha for a given mechanical output. More specifically, if the mechanical output of the devices were normalized at the cochlear level the improvement in gain headroom with the BCI versus the Baha were in the range of 10 to 30 dB. Conclusion: Using a BCI, significantly higher gain setting can be used without feedback problems as compared with using a PBCD.
Ämnesord
- TEKNIK OCH TEKNOLOGIER -- Elektroteknik och elektronik -- Reglerteknik (hsv//swe)
- ENGINEERING AND TECHNOLOGY -- Electrical Engineering, Electronic Engineering, Information Engineering -- Control Engineering (hsv//eng)
Nyckelord
- Baha
- Bone-conduction implant
- Feedback
- Gain headroom
- Implantable transducer
- Loop gain
- Percutaneous bone-conduction device
- Sound radiation
- Stability
- Gain headroom