Evaluation of the impact of a system for real-time visualisation of occupational radiation dose rate during fluoroscopically guided procedures

Sandblom, Viktor, 1987 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiofysik,Institute of Clinical Sciences, Department of Radiation Physics
Mai, Tu, 1987 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiofysik,Institute of Clinical Sciences, Department of Radiation Physics
Almén, Anja, 1964 (författare)
visa fler...
Rystedt, Hans, 1951 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för pedagogik, kommunikation och lärande,Department of Education, Communication and Learning
Cederblad, Åke, 1947 (författare)
Båth, Magnus, 1974 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiofysik,Institute of Clinical Sciences, Department of Radiation Physics
Lundh, Charlotta, 1977 (författare)
visa färre...
 (creator_code:org_t)
2013-07-04
2013
Engelska.
Ingår i: Journal of Radiological Protection. - : IOP Publishing. - 0952-4746 .- 1361-6498. ; 33:3, s. 693-702
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Optimisation of radiological protection for operators working with fluoroscopically guided procedures has to be performed during the procedure, under varying and difficult conditions. The aim of the present study was to evaluate the impact of a system for real-time visualisation of radiation dose rate on optimisation of occupational radiological protection in fluoroscopically guided procedures. Individual radiation dose measurements, using a system for real-time visualisation, were performed in a cardiology laboratory for three cardiologists and ten assisting nurses. Radiation doses collected when the radiation dose rates were not displayed to the staff were compared to radiation doses collected when the radiation dose rates were displayed. When the radiation dose rates were displayed to the staff, one cardiologist and the assisting nurses (as a group) significantly reduced their personal radiation doses. The median radiation dose (Hp(10)) per procedure decreased from 68 to 28 μSv (p = 0.003) for this cardiologist and from 4.3 to 2.5 μSv (p = 0.001) for the assisting nurses. The results of the present study indicate that a system for real-time visualisation of radiation dose rate may have a positive impact on optimisation of occupational radiological protection. In particular, this may affect the behaviour of staff members practising inadequate personal radiological protection.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Medicinsk bioteknologi -- Biomedicinsk laboratorievetenskap/teknologi (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Medical Biotechnology -- Biomedical Laboratory Science/Technology (hsv//eng)
MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy