Derivation and application of dose reduction factors for protective eyewear worn in interventional radiology and cardiology.

Magee, Jill S (författare)
Martin, Colin J (författare)
Sandblom, Viktor, 1987 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiofysik,Institute of Clinical Sciences, Department of Radiation Physics
visa fler...
Carter, Matthew J (författare)
Almén, Anja, 1964 (författare)
Cederblad, Åke, 1947 (författare)
Jonasson, Pernilla (författare)
Lundh, Charlotta (författare)
visa färre...
 (creator_code:org_t)
2014-10-21
2014
Engelska.
Ingår i: Journal of radiological protection : official journal of the Society for Radiological Protection. - : IOP Publishing. - 1361-6498. ; 34:4, s. 811-823
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Doses to the eyes of interventional radiologists and cardiologists could exceed the annual limit of 20mSv proposed by the International Commission on Radiological Protection. Lead glasses of various designs are available to provide protection, but standard eye dosemeters will not take account of the protection they provide. The aim of this study has been to derive dose reduction factors (DRFs) equal to the ratio of the dose with no eyewear, divided by that when lead glasses are worn. Thirty sets of protective eyewear have been tested in x-ray fields using anthropomorphic phantoms to simulate the patient and clinician in two centres. The experiments performed have determined DRFs from simulations of interventional procedures by measuring doses to the eyes of the phantom representing the clinician, using TLDs in Glasgow, Scotland and with an electronic dosemeter in Gothenburg, Sweden. During interventional procedures scattered x-rays arising from the patient will be incident on the head of the clinician from below and to the side. DRFs for x-rays incident on the front of lead glasses vary from 5.2 to 7.6, while values for orientations similar to those used in the majority of clinical practice are between 1.4 and 5.2. Specialised designs with lead glass side shields or of a wraparound style with angled lenses performed better than lead glasses based on the design of standard spectacles. Results suggest that application of a DRF of 2 would provide a conservative factor that could be applied to personal dosemeter measurements to account for the dose reduction provided by any type of lead glasses provided certain criteria relating to design and consistency of use are applied.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy