Bone response to surface modified titanium implants - studies on the tissue response after 1 year to machined and electropolished implants with different oxide thicknesses.

Larsson Wexell, Cecilia, 1965 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för de kirurgiska disciplinerna, Avdelningen för biomaterialvetenskap,Institute of Surgical Sciences, Department of Biomaterials
Emanuelsson, Lena, 1961 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för de kirurgiska disciplinerna, Avdelningen för biomaterialvetenskap,Institute of Surgical Sciences, Department of Biomaterials
Thomsen, Peter, 1953 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för de kirurgiska disciplinerna, Avdelningen för biomaterialvetenskap,Institute of Surgical Sciences, Department of Biomaterials
visa fler...
Ericson, L E (författare)
Aronsson, B O (författare)
Kasemo, B (författare)
Lausmaa, Jukka (författare)
Gothenburg University,Göteborgs universitet,Institutionen för de kirurgiska disciplinerna, Avdelningen för biomaterialvetenskap,Institute of Surgical Sciences, Department of Biomaterials
visa färre...
 (creator_code:org_t)
1997
1997
Engelska.
Ingår i: Journal of materials science. Materials in medicine. - 0957-4530. ; 8:12, s. 721-9
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • The bone formation around titanium implants with varied surface properties was investigated after 1 year in rabbits. Machined and electropolished samples with and without thick, anodically formed surface oxides were prepared, surface characterized and inserted in the cortical bone of rabbits. Scanning electron microscopy, scanning Auger electron spectroscopy and atomic force microscopy revealed marked differences in oxide thickness, surface topography and roughness, but no significant differences in surface chemical composition between the different groups of implants. Light microscopic morphology and morphometry showed that all implants were in contact with bone and had a large proportion of bone within the threads. There were no significant differences between the differently prepared implant groups. Our study shows that a high degree of bone contact and bone formation is achieved after 1 year with titanium implants which are modified with respect to oxide thickness and surface topography. There is no indication that a reduction of surface roughness, which in the initial phase decreases the rate of bone formation, had any influence on the amount of bone after 1 year in rabbit cortical bone.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Odontologi (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Dentistry (hsv//eng)

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy