Method for determining the two-dimensional presampling modulation transfer function in digital radiography

Båth, Magnus, 1974 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för särskilda specialiteter, Avdelningen för radiofysik,Institute of Selected Clinical Sciences, Department of Radiation Physics
Sund, Patrik (författare)
Gothenburg University,Göteborgs universitet,Institutionen för särskilda specialiteter, Avdelningen för radiofysik,Institute of Selected Clinical Sciences, Department of Radiation Physics
Månsson, Lars Gunnar, 1951 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för särskilda specialiteter, Avdelningen för radiofysik,Institute of Selected Clinical Sciences, Department of Radiation Physics
 (creator_code:org_t)
SPIE, 2001
2001
Engelska.
Ingår i: Proceedings of SPIE - The International Society for Optical Engineering. - : SPIE. - 0277-786X. ; 4320, s. 268-279
  • Tidskriftsartikel (övrigt vetenskapligt/konstnärligt)
Abstract Ämnesord
Stäng  
  • In digital radiography, radial asymmetry may be present. The use of a one-dimensional representation of the resolution properties can therefore be questioned. Although measurements are often done in two orthogonal directions, there may be a need for a more detailed description. A method of measuring the two-dimensional presampling modulation transfer function (MTF) has therefore been developed. A finely sampled "disk spread function" is obtained by imaging an aperture mask, consisting of N2 circular holes arranged in an N×N manner in an opaque material, in such a way that each hole is positioned at a different phase relative to the sampling coordinates of the detector system. This spread function is resampled, extrapolated, Fourier transformed, and finally corrected for the finite hole size in order to obtain the presampling MTF. The method was tested on a computed radiography (CR) system through measurements with a prototype mask, consisting of 100 holes of radius 0.2 mm drilled in a lead alloy. The results were compared with measurements using the slit method, and were found to be consistent. Problems associated with the method, e.g. errors due to incorrect alignment of the holes in the aperture mask with the beam, and limitations due to the finite hole size, are discussed.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Nyckelord

Digital radiography
Image quality
Modulation transfer function
MTF

Publikations- och innehållstyp

vet (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Hitta mer i SwePub

Av författaren/redakt...
Båth, Magnus, 19 ...
Sund, Patrik
Månsson, Lars Gu ...
Om ämnet
MEDICIN OCH HÄLSOVETENSKAP
MEDICIN OCH HÄLS ...
och Klinisk medicin
och Radiologi och bi ...
Artiklar i publikationen
Proceedings of S ...
Av lärosätet
Göteborgs universitet

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy