Impact of dose on observer performance in breast tomosynthesis using breast specimens

Timberg, Pontus (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
Båth, Magnus, 1974 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiofysik,Institute of Clinical Sciences, Department of Radiation Physics
Andersson, Ingvar (författare)
Lund University,Lunds universitet,Institutionen för translationell medicin,Medicinska fakulteten,Department of Translational Medicine,Faculty of Medicine
visa fler...
Svahn, Tony (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
Ruschin, Mark (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
Hemdal, Bengt (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
Mattsson, S. (författare)
Tingberg, Anders (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
visa färre...
 (creator_code:org_t)
SPIE, 2008
2008
Engelska.
Ingår i: Progress in Biomedical Optics and Imaging - Proceedings of SPIE. - : SPIE. - 1605-7422. ; 6913, s. 9134-9134
  • Tidskriftsartikel (övrigt vetenskapligt/konstnärligt)
Abstract Ämnesord
Stäng  
  • The purpose of this study was to investigate the effect of dose on lesion detection† and characterization in breast tomosynthesis (BT), using human breast specimens. Images of 27 lesions in breast specimens were acquired on a BT prototype‡ based on a Mammomat Novation (Siemens) full-field digital mammography (FFDM) system. Two detector modes - binned (2×1 in the scan direction) and full resolution - and four BT exposure levels - approximately 2×, 1.5×, 1×, and 0.5× the total mAs at the same beam quality as used in a single FFDM view with a Mammomat Novation unit under automatic exposure control (AEC) conditions - were examined. The exposure for all BT scans was equally divided among 25 projections. An enhanced filtered back projection reconstruction method was applied with a constant filter setting. A human observer performance study was conducted in which the observers were forced to select the minimum (threshold) exposure level at which each lesion could be both detected and characterized for assessment of recall or not in a screening situation. The median threshold exposure level for all observers and all lesions corresponded to approximately 1×, which is half the exposure of what we currently use for BT. A substantial variation in exposure thresholds was noticed for different lesion types. For low contrast lesions with diffuse borders, an exposure threshold of approximately 2× was required, whereas for spiculated high contrast lesions and lesions with well defined borders, the exposure threshold was lower than 0.5×. The use of binned mode had no statistically significant impact on observer performance compared to full resolution mode. There was no substantial difference between the modes for the detection and characterization of the lesion types.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Publikations- och innehållstyp

vet (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy