Comparison of two methods for evaluation of image quality of lumbar spine radiographs

Tingberg, Anders (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
Båth, Magnus, 1974 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för särskilda specialiteter, Avdelningen för radiofysik,Institute of Selected Clinical Sciences, Department of Radiation Physics
Håkansson, Markus (författare)
Gothenburg University,Göteborgs universitet,Institutionen för särskilda specialiteter, Avdelningen för radiofysik,Institute of Selected Clinical Sciences, Department of Radiation Physics
visa fler...
Medin, Joakim (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
Sandborg, M. (författare)
Alm-Carlsson, G. (författare)
Mattsson, Sören (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
Månsson, Lars Gunnar, 1951 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för särskilda specialiteter, Avdelningen för radiofysik,Institute of Selected Clinical Sciences, Department of Radiation Physics
visa färre...
 (creator_code:org_t)
SPIE, 2004
2004
Engelska.
Ingår i: Progress in Biomedical Optics and Imaging - Proceedings of SPIE. - : SPIE. - 1605-7422. ; 5372, s. 251-262
  • Tidskriftsartikel (övrigt vetenskapligt/konstnärligt)
Abstract Ämnesord
Stäng  
  • Purpose To evaluate the image quality of clinical radiographs with two different methods, and to find correlations between the two methods. Methods Based on fifteen lumbar spine radiographs, two new sets of images were created. A hybrid image set was created by adding two distributions of artificial lesions to each original image. The image quality parameters spatial resolution and noise were manipulated and a total of 210 hybrid images were created. A set of 105 disease-free images was created by applying the same combinations of spatial resolution and noise to the original images. The hybrid images were evaluated with the free-response forced error experiment (FFE) and the normal images with visual grading analysis (VGA) by nine experienced radiologists. Results The VGA study showed that images with low noise are preferred over images with higher noise levels. The alteration of the MTF had a limited influence on the VGA score. For the FFE study the visibility of the lesions was independent of the spatial resolution and the noise level. Discussion In this study we found no correlation between the two methods, probably because the detectability of the artificial lesions was not influenced by the manipulations of noise level and resolution. Hence, the detection of lesions in lumbar spine radiography may not be a quantum-noise limited task. The results show the strength of the VGA technique in terms of detecting small changes in the two image quality parameters. The method is more robust and has a higher statistical power than the ROC related method and could therefore, in some cases, be more suitable for use in optimization studies.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Nyckelord

Added pathology
Diagnostic radiology
European Image Criteria
Evaluation methods for clinical image quality
Free-response forced error (FFE) experiment
Lumbar spine hybrid images
Lumbar spine radiography
Visual grading analysis (VGA)

Publikations- och innehållstyp

vet (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy