Evaluation of a corrected implementation of a method of simulating pulmonary nodules in chest tomosynthesis.

Svensson, Frida (författare)
Söderman, Christina (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiofysik,Institute of Clinical Sciences, Department of Radiation Physics
Svalkvist, Angelica (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för biomaterialvetenskap,Institute of Clinical Sciences, Department of Biomaterials
visa fler...
Rossi-Norrlund, Rauni, 1962 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiologi,Institute of Clinical Sciences, Department of Radiology
Vikgren, Jenny, 1957 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiologi,Institute of Clinical Sciences, Department of Radiology
Johnsson, Åse (Allansdotter), 1966 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiologi,Institute of Clinical Sciences, Department of Radiology
Båth, Magnus, 1974 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiologi,Institute of Clinical Sciences, Department of Radiology
visa färre...
 (creator_code:org_t)
2016-09-30
2017
Engelska.
Ingår i: Acta radiologica. - : SAGE Publications. - 1600-0455 .- 0284-1851. ; 58:4, s. 408-413
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • A method of simulating pulmonary nodules in tomosynthesis images has previously been developed and evaluated. An unknown feature of a rounding function included in the computer code was later found to introduce an artifact, affecting simulated nodules in low-signal regions of the images. The computer code has now been corrected.To perform a thorough evaluation of the corrected nodule-simulation method, comparing the detection rate and visual appearance of artificial nodules with those of real nodules in an observer performance experiment.A cohort of 64 patients with a total of 129 pulmonary nodules was used in the study. Artificial nodules, each matching a corresponding real nodule by size, attenuation, and anatomical location, were generated and simulated into the tomosynthesis images of the different patients. The detection rate and visual appearance of artificial nodules generated using both the corrected and uncorrected computer code were compared to those of real nodules. The results were evaluated using modified receiver operating characteristic (ROC) analyses.The difference in detection rate between artificial and real nodules slightly increased using the corrected computer code (uncorrected code: area under the curve [AUC], 0.47; 95% CI, 0.43-0.51; corrected code: AUC, 0.42; 95% CI, 0.38-0.46). The visual appearance was however substantially improved using the corrected computer code (uncorrected code: AUC, 0.70; 95% CI, 0.63-0.76; corrected code: AUC, 0.49; 95% CI, 0.29-0.65).The computer code including a correct rounding function generates simulated nodules that are more visually realistic than simulated nodules generated using the uncorrected computer code, but have a slightly different detection rate compared to real nodules.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine (hsv//eng)

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy