Detection and Characterization of Solid Pulmonary Nodules at Digital Chest Tomosynthesis: Data from a Cohort of the Pilot Swedish Cardiopulmonary Bioimage Study

Meltzer, Carin (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiologi,Institute of Clinical Sciences, Department of Radiology
Vikgren, Jenny, 1957 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiologi,Institute of Clinical Sciences, Department of Radiology
Bergman, Bengt, 1953 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för medicin,Institute of Medicine
visa fler...
Molnar, David (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiologi,Institute of Clinical Sciences, Department of Radiology
Rossi-Norrlund, Rauni, 1962 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiologi,Institute of Clinical Sciences, Department of Radiology
Hassoun, Asmaa (författare)
Gottfridsson, Bengt (författare)
Båth, Magnus, 1974 (författare)
Johnsson, Åse (Allansdotter), 1966 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiologi,Institute of Clinical Sciences, Department of Radiology
visa färre...
 (creator_code:org_t)
Radiological Society of North America (RSNA), 2018
2018
Engelska.
Ingår i: Radiology. - : Radiological Society of North America (RSNA). - 0033-8419 .- 1527-1315. ; 287:3, s. 1018-1027
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Purpose: To investigate the performance of digital tomosynthesis (DTS) for detection and characterization of incidental solid lung nodules. Materials and Methods: This prospective study was based on a population study with 1111 randomly selected participants (age range, 50-64 years) who underwent a medical evaluation that included chest computed tomography (CT). Among these, 125 participants with incidental nodules 5 mm or larger were included in this study, which added DTS in conjunction with the follow-up CT and was performed between March 2012 and October 2014. DTS images were assessed by four thoracic radiologists blinded to the true number of nodules in two separate sessions according to the 5-mm (125 participants) and 6-mm (55 participants) cut-off for follow-up of incidental nodules. Pulmonary nodules were directly marked on the images by the readers and graded regarding confidence of presence and recommendation for follow-up. Statistical analyses included jackknife free-response receiver operating characteristic, receiver operating characteristic, and Cohen k coefficient. Results: Overall detection rate ranges of CT-proven nodules 5 mm or larger and 6 mm or larger were, respectively, 49%58% and 48%-62%. Jackknife free-response receiver operating characteristics figure of merit for detection of CT-proven nodules 5 mm or larger and 6 mm or larger was 0.47 and 0.51, respectively, and area under the receiver operating characteristic curve regarding recommendation for follow-up was 0.62 and 0.65, respectively. Conclusion: Routine use of DTS would result in lower detection rates and reduced number of small nodules recommended for follow-up. (C) RSNA, 2018.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Nyckelord

observer performance
computed-tomography
radiography
management
radiology
software
viewdex
lesions
tool
Radiology
Nuclear Medicine & Medical Imaging

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

  • Radiology (Sök värdpublikationen i LIBRIS)

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy