Interleukin-6 measured using the automated electrochemiluminescence immunoassay method for the identification of intra-amniotic inflammation in preterm prelabor rupture of membranes.

Musilova, Ivana (author)
Andrys, Ctirad (author)
Holeckova, Magdalena (author)
show more...
Kolarova, Veronika (author)
Pliskova, Lenka (author)
Drahosova, Marcela (author)
Bolehovska, Radka (author)
Pilka, Radovan (author)
Huml, Karel (author)
Cobo, Teresa (author)
Jacobsson, Bo, 1960 (author)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för obstetrik och gynekologi,Institute of Clinical Sciences, Department of Obstetrics and Gynecology
Kacerovsky, Marian (author)
show less...
 (creator_code:org_t)
2018-12-17
2020
English.
In: The journal of maternal-fetal & neonatal medicine : the official journal of the European Association of Perinatal Medicine, the Federation of Asia and Oceania Perinatal Societies, the International Society of Perinatal Obstetricians. - : Informa UK Limited. - 1476-4954. ; 33:11, s. 1919-1926
  • Journal article (peer-reviewed)
Abstract Subject headings
Close  
  • We aimed to compare the amniotic fluid interleukin (IL)-6 concentrations measured using the automated electrochemiluminescence immunoassay method and ELISA, and to establish an IL-6 concentration cut-off value for intra-amniotic inflammation (IAI) in preterm prelabor rupture of membranes (PPROM), which can be used in the automated electrochemiluminescence immunoassay method.A total of 120 women with PPROM were included in this study. Amniotic fluid samples were obtained via transabdominal amniocentesis. IL-6 concentrations were assessed using both the automated electrochemiluminescence immunoassay method and ELISA, the current gold standard. IAI was defined as an amniotic fluid IL-6 concentration of ≥ 2600 pg/mL measured using ELISA.A correlation between both assays was found (Spearman's rho = 0.97; p < 0.0001). Based on the receiver-operating characteristic curve for the identification of IAI (area under the curve = 0.99), a cut-off value of ≥ 3000pg/mL was selected for the automated electrochemiluminescence immunoassay method with a sensitivity of 88%, specificity of 99%, positive predictive value of 97%, negative predictive value of 96%, and likelihood ratio of 76.

Subject headings

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine (hsv//eng)
MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Reproduktionsmedicin och gynekologi (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Obstetrics, Gynaecology and Reproductive Medicine (hsv//eng)

Publication and Content Type

ref (subject category)
art (subject category)

Find in a library

To the university's database

  • 1 of 1
  • Previous record
  • Next record
  •    To hitlist

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Close

Copy and save the link in order to return to this view