A Monte Carlo study of the energy dependence of Al2O3:C crystals for real-time in vivo dosimetry in mammography

Aznar, Marianne C. (författare)
Medin, Johan (författare)
Hemdal, Bengt (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
visa fler...
Thilander-Klang, Anne, 1959 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för särskilda specialiteter, Avdelningen för radiofysik,Extern,Institute of Selected Clinical Sciences, Department of Radiation Physics,External
Bøtter-Jensen, L. (författare)
Mattsson, Sören (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
Medin, Joakim (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
visa färre...
 (creator_code:org_t)
2005-05-17
2005
Engelska.
Ingår i: Radiation Protection Dosimetry. - : Oxford University Press (OUP). - 0144-8420 .- 1742-3406. ; 114:Nos 1-3, s. 444-449
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • In a previous experimental study, a novel method for in vivo dosimetry has been investigated, based on radioluminescence (RL) and optically stimulated luminescence (OSL). However, because of the large difference in atomic composition between the detector material and the breast tissue, relatively large energy dependence in low-energy X-ray beams can be expected. In the present work, the energy dependence of Al2O3:C crystals was modelled with the Monte Carlo code EGSnrc using three types of X-ray spectra. The results obtained (5.6-7.3%) agree with a previously determined experimental result (9%) within the combined standard uncertainty of the two methods. The influence of the size of the crystal on the energy dependence was investigated together with the effect of varying the thickness of the surrounding light-protective material. The results obtained indicate a minor effect owing to the thickness of the light-protective material, and a somewhat larger effect from reducing the diameter of the crystal. The outcome of this study can be used to improve the future design of the RL/OSL dosimetry system for use in mammography.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy