Influence of anode/filter material and tube potential on contrast, signal-to-noise ratio and average absorbed dose in mammography: a Monte Carlo study
-
Dance, David (författare)
-
- Thilander-Klang, Anne, 1959 (författare)
- Gothenburg University,Göteborgs universitet,Institutionen för särskilda specialiteter, Avdelningen för radiofysik,Institute of Selected Clinical Sciences, Department of Radiation Physics
-
Sandborg, Mikael (författare)
-
visa fler...
-
Skinner, Claire (författare)
-
Castellano-Smith, Isabelle (författare)
-
Alm Carlsson, Gudrun (författare)
-
visa färre...
-
(creator_code:org_t)
- 2000
- 2000
- Engelska.
-
Ingår i: British Journal of Radiology. - 0007-1285. ; 73, s. 1056-1067
- Relaterad länk:
-
https://gup.ub.gu.se...
Abstract
Ämnesord
Stäng
- The comparative performance of mammographic X-ray systems that use different anode/filter combinations has been assessed for screen±®lm and digital imaging. Monte Carlo techniques have been used to calculate average glandular dose as well as contrast and signal-tonoise ratio for imaging two test details. Five anode/filter combinations have been studied to establish the potential for dose saving or image quality improvement. For screen±film mammography, it was found that little bene®t is gained by changing from a standard 28 kV molybdenum/molybdenum spectrum for breasts up to 6 cm thick. For thicker breasts, where the tube potential for the standard technique might be increased, 20% improvement in contrast can be achieved without dose penalty using molybdenum/rhodium or rhodium/rhodium spectra, whereas dose savings of more than 50% can be attained whilst maintaining contrast using tungsten/rhodium or rhodium/aluminium spectra. In digital mammography, a molybdenum/ molybdenum spectrum delivers the lowest dose for a 2 cm breast, but gives the highest dose for thicker breasts. Tungsten/rhodium or rhodium/aluminium spectra provide the lowest doses at greater thicknesses. It is concluded that for screen±film mammography, molybdenum/ molybdenum is the spectrum of choice for all but the thickest or most glandular breasts. In digital mammography, an alternative spectrum is preferable for breasts thicker than 2 cm.
Ämnesord
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)
Nyckelord
- Mammography Radaition Dosimetry
Publikations- och innehållstyp
- ref (ämneskategori)
- art (ämneskategori)
Hitta via bibliotek
Till lärosätets databas