Trueness and precision of 5 intraoral scanners for scanning edentulous and dentate complete-arch mandibular casts: A comparative in vitro study

Braian, Michael (författare)
Malmö universitet,Odontologiska fakulteten (OD)
Wennerberg, Ann, 1955 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för odontologi,Institute of Odontology
 (creator_code:org_t)
Elsevier BV, 2019
2019
Engelska.
Ingår i: Journal of Prosthetic Dentistry. - : Elsevier BV. - 0022-3913 .- 1097-6841. ; 122:2, s. 129-136
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Statement of problem. Limited information is available on the trueness and precision of intraoral scanners (IOSs) for scanning dentate and edentulous casts. Purpose. The purpose of this in vitro study was to evaluate the trueness and precision of 5 different IOS devices for scanning a dentate and an edentulous cast in a standardized way for short arches and complete arches. Material and methods. Five IOS devices were used to scan 2 computer metric measured casts using a coordinate measuring machine (CMM). Both were scanned 15 times. All scans were carried out by 1 experienced operator in a standardized way. One cast was edentulous, and 1 was dentate. Five cylindrical landmarks were added to each cast. These cylinders made the measurement of point-to-point distances possible, dividing the tests into cross-arch measurements and intercylindrical (short-arch) measurements. The Student t test, Mann-Whitney test, and Levene test for equality were used to calculate the difference between the edentulous and dentate scans for both cross-arch and intercylindrical measurements (alpha=.05). Results. For the cross-arch measurements on the edentulous scans, the trueness values ranged between 6 mu m (Emerald P1-P2) and 193 mu m (Omnicam P1-P5) and for the intercylindrical measurements, between 2 mu m (Itero P4-P5) and -103 mu m (CS 3600 P1-P2). For the dentate cast, the cross-arch trueness values ranged between 6 mu m (CS 3600 P1-P2) and 150 mu m (TRIOS 3 P1-P5) and for the intercylindrical measurements, between 4 mu m (Itero P4-P5) and -56 mu m (Emerald P4-P5). Conclusions. Significant differences were found in scanning edentulous and dentate scans for short arches and complete arches. Trueness for complete-arch scans were <193 mu m for edentulous scans and <150 mu m for dentate scans. Trueness for short-arch scans were <103 mu m for edentulous scans and <56 mu m for dentate scans.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Odontologi (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Dentistry (hsv//eng)

Nyckelord

accuracy
fit
impressions
silicone
implants
Dentistry
Oral Surgery & Medicine
Dentistry

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Hitta mer i SwePub

Av författaren/redakt...
Braian, Michael
Wennerberg, Ann, ...
Om ämnet
MEDICIN OCH HÄLSOVETENSKAP
MEDICIN OCH HÄLS ...
och Klinisk medicin
och Odontologi
Artiklar i publikationen
Journal of Prost ...
Av lärosätet
Göteborgs universitet
Malmö universitet

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy