Long-Term Outcomes of the Minimally Invasive Ponto Surgery vs. Linear Incision Technique With Soft Tissue Preservation for Installation of Percutaneous Bone Conduction Devices

Strijbos, Ruben M. (författare)
Straatman, Louise V. (författare)
Calon, Tim G.A. (författare)
visa fler...
Johansson, Martin L (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för biomaterialvetenskap,Institute of Clinical Sciences, Department of Biomaterials
de Bruijn, Arthur J.G. (författare)
van den Berge, Herbert (författare)
Wagenaar, Mariette (författare)
Eichhorn, Edwin (författare)
Janssen, Miranda (författare)
Jonhede, Sofia (författare)
van Tongeren, Joost (författare)
Holmberg, Marcus (författare)
Stokroos, Robert (författare)
visa färre...
 (creator_code:org_t)
2021-02-24
2021
Engelska.
Ingår i: Frontiers in Neurology. - : Frontiers Media SA. - 1664-2295. ; 12:112
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Objective: Comparing the surgical outcomes of the Minimally Invasive Ponto Surgery (MIPS) technique with the linear incision technique with soft tissue preservation (LITT-P) for bone conduction devices after a follow-up of 22 months. Methods: In this multicenter randomized controlled trial, there was the inclusion of 64 adult patients eligible for unilateral surgery. There was 1:1 randomization to the MIPS (test) or the LITT-P (control) group. The primary outcome was an (adverse) soft tissue reaction. Secondary outcomes were pain, loss of sensibility, soft tissue height/overgrowth, skin sagging, implant loss, Implant Stability Quotient measurements, cosmetic scores, and quality of life questionnaires. Results: Sixty-three subjects were analyzed in the intention-to-treat population. No differences were found in the presence of (adverse) soft tissue reactions during complete follow-up. Also, there were no differences in pain, wound dehiscence, skin level, soft tissue overgrowth, and overall quality of life. Loss of sensibility (until 3-month post-surgery), cosmetic scores, and skin sagging outcomes were better in the MIPS group. The Implant Stability Quotient was higher after the LITT-P for different abutment lengths at various points of follow-up. Implant extrusion was nonsignificantly higher after the MIPS (15.2%) compared with LITT-P (3.3%). Conclusion: The long-term results show favorable outcomes for both techniques. The MIPS is a promising technique with some benefits over the LITT-P. Concerns regarding nonsignificantly higher implant loss may be overcome with future developments and research. Clinical Trial Registration: www.ClinicalTrials.gov, identifier: NCT02438618.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Medicinsk bioteknologi -- Biomaterialvetenskap (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Medical Biotechnology -- Biomaterials Science (hsv//eng)
MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Oto-rhino-laryngologi (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Otorhinolaryngology (hsv//eng)

Nyckelord

bone conduction device (BCD)
hearing loss
minimally invasive ponto surgery
MIPS
soft tissue reactions
surgical outcomes
surgical technique
tissue preservation

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy