Investigation of the Impact of Defective Ultrasound Transducers on Clinical Image Quality in Grayscale 2-D Still Images.

Lorentsson, Robert, 1969 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för medicinsk strålningsvetenskap,Institute of Clinical Sciences, Department of Medical Radiation Sciences
Hosseini, Nasser (författare)
Aurell, Ylva, 1955 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiologi,Institute of Clinical Sciences, Department of Radiology
visa fler...
Collin, David (författare)
Frösing, Eva (författare)
Szaro, Pavel, 1981 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiologi,Institute of Clinical Sciences, Department of Radiology
Månsson, Lars Gunnar, 1951 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för medicinsk strålningsvetenskap,Institute of Clinical Sciences, Department of Medical Radiation Sciences
Båth, Magnus, 1974 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för medicinsk strålningsvetenskap,Institute of Clinical Sciences, Department of Medical Radiation Sciences
visa färre...
 (creator_code:org_t)
2023
2023
Engelska.
Ingår i: Ultrasound in medicine & biology. - 0301-5629 .- 1879-291X. ; 49:9, s. 2126-2133
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • There are several studies that show high defect rates of transducers in clinical use. The purpose of the present study was to investigate whether image quality and the risk for misdiagnosis is affected by using defective transducers.Four defective transducers with varying degrees of defect severity, still in clinical use, were selected. Forty artifact-affected clinical images from each transducer were compared with images acquired from fully functional transducers, of the same model, in an observer study where four experienced radiologists rated each of the 320 images. The rating tasks included if the artifacts were detectable, if the possible artifacts might affect the diagnosis, how well structural details were reproduced and, finally, an assessment of overall image quality.The artifacts in the images were detectable for three of the four transducers (p < 0.05), and in 121 of 640 assessments of the images from the defective transducers the observers were confident that the artifacts could affect the diagnosis. All four faulty transducers were assessed to have decreased ability to resolve structural details (p < 0.05), and three of the four transducers were assessed to have worse overall image quality (p < 0.05).The present study shows that image quality and the risk of misdiagnosis can be affected by using defective transducers. This highlights the importance of frequent quality control of the transducers to avoid decreased image quality and even misdiagnosis.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Nyckelord

Artifacts
Quality Control
Transducers

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy