Hydroxyapatite enhances long-term fixation of titanium implants

Reikeras, O. (författare)
Johansson, Carina B. (författare)
Örebro universitet,Institutionen för teknik
Sundfeldt, Mikael, 1962 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper,Institute of Clinical Sciences
 (creator_code:org_t)
Boca Raton, FL : CRC Press, 2006
2006
Engelska.
Ingår i: J Long Term Eff Med Implants. - Boca Raton, FL : CRC Press. - 1050-6934 .- 1940-4379. ; 16:2, s. 165-73
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • The aim of this study was to evaluate osseous integration of hydroxyapatite coated titanium implants over time as compared to pure titanium. In 20 rats the medullary cavity of both femoral bones was entered by an awl from the trochanteric area. With steel burrs it was successively reamed to a diameter of 2.0 mm. Nails with a diameter of 2.0 mm and with a length of 34 mm were inserted into the medullary cavity; a pure titanium nail on the left side and a titanium nail entirely plasma-sprayed with hydroxyapatite (HA) on the right side. The surface roughness of the pure titanium was characterized by Ra 2.6 microm and Rt 22 microm, and HA had a roughness of Ra 7.5 (arithmetical mean roughness) microm and Rt (maximum profile height) 52 microm. The rats were randomized to a follow-up of 6 and 12 months, respectively. At sacrifice the femoral bones were dissected free from soft tissues. The bones were radiographed and then immersed in fixative. A specimen-slice of about 5 mm thickness was prepared from the region under the trochanter minor with a water cooled band-saw. Sample preparation for undecalcified tissue followed the internal guidelines at the laboratories of Biomaterials/Handicap Research. At 6 months the median bone bonding contact of the implants was 40% (range 0-92) in the titanium group and 34% (0-86) in the HA group. At 12 months the median bone bonding contact was 51% (0-97) in the titanium group and 86% (72-98) in the HA group. In conclusion, we found a significant (p = 0.001) increase in bone bonding contact from 6 to 12 months of the HA coated nails and significantly (p = 0.043) enhanced bone bonding contact in HA coated nails at 12 months as compared to pure titanium nails.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Kirurgi (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Surgery (hsv//eng)

Nyckelord

Animals
Biocompatible Materials/metabolism
Durapatite/*metabolism
Femur/cytology/*metabolism/*surgery
*Internal Fixators
Male
Osseointegration/physiology
*Prostheses and Implants
Prosthesis Implantation
Rats
Rats
Wistar
Time Factors
Titanium/*metabolism
MEDICINE
Medicine

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Hitta mer i SwePub

Av författaren/redakt...
Reikeras, O.
Johansson, Carin ...
Sundfeldt, Mikae ...
Om ämnet
MEDICIN OCH HÄLSOVETENSKAP
MEDICIN OCH HÄLS ...
och Klinisk medicin
och Kirurgi
Artiklar i publikationen
J Long Term Eff ...
Av lärosätet
Göteborgs universitet
Örebro universitet

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy