Testing the implantable cardioverter-defibrillator after implantation--is it necessary?

Sandstedt, Bengt, 1951 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för medicin, avdelningen för akut och kardiovaskulär medicin,Institute of Medicine, Department of Emergeny and Cardiovascular Medicine
Gottfridsson, Christer, 1958 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för medicin, avdelningen för molekylär och klinisk medicin,Institute of Medicine, Department of Molecular and Clinical Medicine
Nyström, Britta, 1946 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för medicin, avdelningen för molekylär och klinisk medicin,Institute of Medicine, Department of Molecular and Clinical Medicine
visa fler...
Edvardsson, Nils, 1942 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för medicin, avdelningen för molekylär och klinisk medicin,Institute of Medicine, Department of Molecular and Clinical Medicine
visa färre...
 (creator_code:org_t)
2007
2007
Engelska.
Ingår i: Pacing Clin Electrophysiol. - 0147-8389. ; 30:8, s. 985-91
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • The results of intraoperative and postoperative predischarge implantable cardioverter-defibrillator (ICD) testing of 211 consecutive patients, starting at 15 J and requiring two successful terminations of induced VT/VF with a relative defibrillation safety margin (DSM) of >10 J, were reviewed. The aim was to define the type of intraoperative response that would make postoperative predischarge testing unnecessary. The intraoperative responses were divided into three types: A, a DSM > or =10 J and an absolute energy level of < or =20 J; B, a DSM of > or =10 J and an absolute energy level of >20 J; and C, a DSM <10 J and an absolute energy level of >20 J. At operation, the responses to defibrillation were A, 88.6%; B, 7.1%; and C, 4.3%. Accepting an A response only would leave 11.4% of the patients for postoperative testing. The positive and negative predictive values for diagnosing a postoperative C response were 0.78 and 0.97, respectively. Similarly, the predictive values for diagnosing a postoperative B or C response were 0.71 and 0.97, respectively. The postoperative testing responses were A, 89.1%; B, 4.3%; and C, 6.6%. In summary, an intraoperative A response was sufficient to make a postoperative defibrillation testing unnecessary, while it was found that intraoperative B and C responders should undergo postoperative testing. Applying these criteria, approximately 90% of the patients could be discharged without any postoperative induction test.

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy