Accuracy of proximal caries depth measurements: comparison between limited cone beam computed tomography, storage phosphor and film radiography

Akdeniz, BG (författare)
Gröndahl, Hans-Göran, 1940 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för odontologi,Institute of Odontology
Magnusson, Bengt, 1942 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för odontologi,Institute of Odontology
 (creator_code:org_t)
2006
2006
Engelska.
Ingår i: Caries Res. ; 40:3, s. 202-207
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • The aim of this study was to compare the accuracy of limited cone beam computed tomography (LCBCT), an image plate system and F-speed film in assessing the depth of proximal carious lesions. Radiographs of a dry mandible with sound and carious teeth were obtained with all three methods. In 41 molar and premolar proximal surfaces, 2 observers independently measured lesion depth on the images from the three modalities. The correlation of measurements was assessed with Pearson's correlation analysis. Results from imaging modalities and histological sectioning (gold standard) were compared using Bland-Altman plots. Overall comparison of the depth measurements from the imaging modalities and the gold standard was done using repeated-measures ANOVA. Pairwise comparisons of systems were done by the Bonferroni t test. The correlation between the measurements of the two observers was 0.977 for film, 0.997 for image plate system and 0.998 for LCBCT. Bland-Altman plots revealed that LCBCT agreed very closely with the gold standard while the agreement between the latter and the image plate--or film--images was moderate. The mean difference and 95% limits of agreement between LCBCT and the gold standard were smaller than those between either image plate or F-speed film and the gold standard. The LCBCT method appears as a promising tool for detection and monitoring of proximal carious lesions.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Odontologi (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Dentistry (hsv//eng)

Nyckelord

Cone beam X-ray CT
digital radiography

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Hitta mer i SwePub

Av författaren/redakt...
Akdeniz, BG
Gröndahl, Hans-G ...
Magnusson, Bengt ...
Om ämnet
MEDICIN OCH HÄLSOVETENSKAP
MEDICIN OCH HÄLS ...
och Klinisk medicin
och Odontologi
Artiklar i publikationen
Av lärosätet
Göteborgs universitet

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy