First clinical experience of a ring-configured cadmium zinc telluride camera : A comparative study versus conventional gamma camera systems

Cerić Andelius, Irma (författare)
Skåne University Hospital
Minarik, David (författare)
Lund University,Lunds universitet,Klinisk fysiologi och nuklearmedicin, Malmö,Forskargrupper vid Lunds universitet,Clinical Physiology and Nuclear Medicine, Malmö,Lund University Research Groups,Skåne University Hospital,Wallenberg Centre for Molecular Medicine (WCMM)
Persson, Eva (författare)
Lund University,Lunds universitet,Klinisk fysiologi, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Clinical Physiology (Lund),Section V,Department of Clinical Sciences, Lund,Faculty of Medicine,Skåne University Hospital
visa fler...
Mosén, Henrik (författare)
Lund University,Lunds universitet,Klinisk fysiologi, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Clinical Physiology (Lund),Section V,Department of Clinical Sciences, Lund,Faculty of Medicine,Skåne University Hospital
Valind, Kristian (författare)
Lund University,Lunds universitet,Klinisk fysiologi och nuklearmedicin, Malmö,Forskargrupper vid Lunds universitet,WCMM- Wallenberg center för molekylär medicinsk forskning,Medicinska fakulteten,Clinical Physiology and Nuclear Medicine, Malmö,Lund University Research Groups,WCMM-Wallenberg Centre for Molecular Medicine,Faculty of Medicine,Skåne University Hospital
Trägårdh, Elin (författare)
Lund University,Lunds universitet,Nuklearmedicin, Malmö,Forskargrupper vid Lunds universitet,Klinisk fysiologi och nuklearmedicin, Malmö,WCMM- Wallenberg center för molekylär medicinsk forskning,Medicinska fakulteten,Nuclear medicine, Malmö,Lund University Research Groups,Clinical Physiology and Nuclear Medicine, Malmö,WCMM-Wallenberg Centre for Molecular Medicine,Faculty of Medicine,Skåne University Hospital
Oddstig, Jenny (författare)
Lund University,Lunds universitet,Klinisk fysiologi och nuklearmedicin, Malmö,Forskargrupper vid Lunds universitet,WCMM- Wallenberg center för molekylär medicinsk forskning,Medicinska fakulteten,Clinical Physiology and Nuclear Medicine, Malmö,Lund University Research Groups,WCMM-Wallenberg Centre for Molecular Medicine,Faculty of Medicine,Skåne University Hospital
visa färre...
 (creator_code:org_t)
2024
2024
Engelska 10 s.
Ingår i: Clinical Physiology and Functional Imaging. - 1475-0961. ; 44:1, s. 79-88
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • BACKGROUND: A novel semiconductor cadmium zinc telluride (CZT) gamma camera system using a block sequential regularized expectation maximization (BSREM) reconstruction algorithm is now clinically available. Here we investigate how a multi-purpose ring-configurated CZT system can be safely applied in clinics and describe the initial optimization process.METHOD: Seventy-six patients (bone-, cardiac- and lung scan) were scanned on a conventional gamma camera (planar and/or single-photon emission computed tomography [SPECT]/SPECT-CT) used in clinical routine and on the ring-configurated CZT camera Starguide (GE Healthcare). These data were used to validate and optimize the Starguide system for routine clinical use.RESULTS: Comparable image quality for the Starguide system, to that of the conventional gamma camera, was achieved for bone scan (4 min/bed position [BP] using a relative difference prior [RDP] with gamma 2 and beta 0.4, along with 10 iterations and 10 subsets), cardiac scan (8 min [stress] and 3 min 20 s [rest] using median root prior [MRP] with beta 0.07 non attenuation corrected and 0.008 attenuation corrected and 50 interations and 10 subsets for both stress and rest) and lung scan (10 min [vent] and 5 min [perf] using RDP with gamma 0.5 and beta 0.03 [vent] and 0.02 [perf] and 20 interations and 10 subsets for both vent and perf).CONCLUSIONS: It was possible to transition from a conventional gamma camera to the Starguide system as part of the clinical routine, with acceptable image quality. Images from the Starguide system were deemed to be at least as good as those from a conventional gamma camera.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)
TEKNIK OCH TEKNOLOGIER  -- Medicinteknik -- Medicinsk bildbehandling (hsv//swe)
ENGINEERING AND TECHNOLOGY  -- Medical Engineering -- Medical Image Processing (hsv//eng)

Publikations- och innehållstyp

art (ämneskategori)
ref (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy