Interstitial laser thermotherapy : Comparison between bare fibre and sapphire probe

Möller, Páll H. (författare)
Lund University,Lunds universitet,Kirurgi, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Medicinsk strålningsfysik, Lund,Institutionen för kliniska vetenskaper, Lund,Surgery (Lund),Section V,Department of Clinical Sciences, Lund,Faculty of Medicine,Medical Radiation Physics, Lund,Department of Clinical Sciences, Lund
Lindberg, Lars (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Medical Radiation Physics, Lund,Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
Henriksson, Pär H. (författare)
Skåne University Hospital
visa fler...
Persson, Bertil R.R. (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Medical Radiation Physics, Lund,Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
Tranberg, Karl G. (författare)
Lund University,Lunds universitet,Kirurgi, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Medicinsk strålningsfysik, Lund,Institutionen för kliniska vetenskaper, Lund,Surgery (Lund),Section V,Department of Clinical Sciences, Lund,Faculty of Medicine,Medical Radiation Physics, Lund,Department of Clinical Sciences, Lund
visa färre...
 (creator_code:org_t)
1995
1995
Engelska.
Ingår i: Lasers in Medical Science. - 0268-8921. ; 10:3, s. 193-200
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • A sapphire probe and a bare fibre were compared with respect to temperature control and distribution and light fluence in interstitial laser thermotherapy. Experiments were performed in processed liver using an Nd-YAG laser and output power levels of 1-4 W. The temperature was controlled at a distance of 10 mm using a feedback circuit with an automatic thermometry system and thermistor probes. With the sapphire probe, carbonization was rare at power levels of 1-2 W but was observed in half of the experiments at 3 W and in all experiments at 4 W. Using the bare fibre, carbonization was seen in almost all experiments. Absence of carbonization was associated with a moderate decrease in the penetration of light and excellent control of the temperature, whereas carbonization led to rapid impairment of light penetration and temperature control. In addition, the temperature gradient was smaller with the sapphire probe than with the bare fibre or when carbonization was absent. It is concluded that a diffuser tip, such as the sapphire probe, may be preferable to the bare fibre for interstitial laser thermotherapy because it gives a smaller temperature gradient and helps to avoid carbonization which results in preserved light penetration and improved temperature control.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Nyckelord

Feedback control
Hyperthermia
Light penetration
Nd-YAG laser
Temperature distribution

Publikations- och innehållstyp

art (ämneskategori)
ref (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy