Allergic contact dermatitis caused by isobornyl acrylate in the Enlite glucose sensor and the Paradigm MiniMed Quick-set insulin infusion set

Herman, Anne (författare)
Saint-Luc University Hospital
Baeck, Marie (författare)
Saint-Luc University Hospital
de Montjoye, Laurence (författare)
Saint-Luc University Hospital
visa fler...
Bruze, Magnus (författare)
Lund University,Lunds universitet,Yrkes- och miljödermatologi,Forskargrupper vid Lunds universitet,Occupational and Environmental Dermatology,Lund University Research Groups,Skåne University Hospital
Giertz, Emil (författare)
Blekinge Hospital
Goossens, An (författare)
University Hospitals Leuven
Mowitz, Martin (författare)
Lund University,Lunds universitet,Yrkes- och miljödermatologi,Forskargrupper vid Lunds universitet,Occupational and Environmental Dermatology,Lund University Research Groups,Skåne University Hospital
visa färre...
 (creator_code:org_t)
2019-08-27
2019
Engelska.
Ingår i: Contact Dermatitis. - : Wiley. - 0105-1873 .- 1600-0536. ; 81:6, s. 432-437
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Background: The FreeStyle Libre glucose sensor has caused many cases of allergic contact dermatitis, and isobornyl acrylate (IBOA) in this sensor has been identified as one of the culprit allergens. Objectives: To report on the presence of IBOA in devices produced by Medtronic, namely, Enlite sensor and insulin infusion set Paradigm MiniMed Quick-set. Patients and Methods: Five patients reacting to the glucose sensor Enlite and/or the insulin infusion set Paradigm MiniMed Quick-set observed in three clinics (two Belgian and one Swedish) were patch tested with the baseline and other series, as well as with IBOA; four of them also with pieces of adhesive patches from the devices, and two with a thin layer chromatogram of Enlite glucose sensor extracts. Gas chromatography-mass spectrometry (GC-MS) analyses were performed. Results: Four patients reacted to IBOA and one to colophonium, a known allergen in Enlite, and three to the adhesive part of the sensor or the insulin infusion set. IBOA was identified in the sensor by GC-MS, and its presence was indicated in the infusion set. Conclusions: IBOA is a contact allergen in Enlite glucose sensor, and likely also in the infusion set. Therefore, these devices are not suitable alternatives for patients sensitized to the FreeStyle Libre sensor.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Dermatologi och venereologi (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Dermatology and Venereal Diseases (hsv//eng)

Nyckelord

acrylates
allergic contact dermatitis
delayed hypersensitivity
glucose sensor
insulin pump
isobornyl acrylate
medical device

Publikations- och innehållstyp

art (ämneskategori)
ref (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy