Sökning: id:"swepub:oai:lup.lub.lu.se:157e120f-2836-4931-bc4a-04a11c653289" > Generation of X-ray...

Generation of X-rays For Medical Imaging By High-power Lasers - Preliminary-results

Herrlin, K (författare)
Svahn, G (författare)
Olsson, C (författare)
visa fler...
Pettersson, H (författare)
Tillman, C (författare)
Persson, Anders (författare)
Lund University,Lunds universitet,Atomfysik,Fysiska institutionen,Institutioner vid LTH,Lunds Tekniska Högskola,Atomic Physics,Department of Physics,Departments at LTH,Faculty of Engineering, LTH
Wahlström, Claes-Göran (författare)
Lund University,Lunds universitet,Atomfysik,Fysiska institutionen,Institutioner vid LTH,Lunds Tekniska Högskola,Atomic Physics,Department of Physics,Departments at LTH,Faculty of Engineering, LTH
Svanberg, Sune (författare)
Lund University,Lunds universitet,Atomfysik,Fysiska institutionen,Institutioner vid LTH,Lunds Tekniska Högskola,Atomic Physics,Department of Physics,Departments at LTH,Faculty of Engineering, LTH
visa färre...
 (creator_code:org_t)
1993
1993
Engelska.
Ingår i: Radiology. - 1527-1315. ; 189:1, s. 65-68
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • PURPOSE: To evaluate the use of x-ray imaging performed with a high-power laser system in biologic and medical studies. MATERIALS AND METHODS: A compact terawatt laser system based on chirped pulse amplification in titanium-doped sapphire was used. X rays were generated by irradiating a high-atomic-number target (tantalum). RESULTS: When photons with an energy below 10 keV were removed with use of 3 mm of aluminum, the half-value layer in aluminum for the remaining x rays was approximately 10 mm. The x-ray source allowed performance of biologic magnification radiography. Star-pattern tests indicated an equivalent focal spot size of less than 60 mum. Exposures of a single pulse could be obtained. The duration of a single x-ray pulse was estimated to be of the order of picoseconds. CONCLUSION. With use of subpicosecond laser pulses, x-ray generation can occur with a smaller equivalent focal spot size than with conventional x-ray sources.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Publikations- och innehållstyp

art (ämneskategori)
ref (ämneskategori)

Hitta via bibliotek

  • Radiology (Sök värdpublikationen i LIBRIS)

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy