Whole-heart four-dimensional flow can be acquired with preserved quality without respiratory gating, facilitating clinical use : A head-to-head comparison

Kanski, Mikael (författare)
Lund University,Lunds universitet,Klinisk fysiologi, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Clinical Physiology (Lund),Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
Töger, Johannes (författare)
Lund University,Lunds universitet,Klinisk fysiologi, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Clinical Physiology (Lund),Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
Steding-Ehrenborg, Katarina (författare)
Lund University,Lunds universitet,Klinisk fysiologi, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Clinical Physiology (Lund),Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
visa fler...
Xanthis, Christos (författare)
Lund University,Lunds universitet,Klinisk fysiologi, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Clinical Physiology (Lund),Section V,Department of Clinical Sciences, Lund,Faculty of Medicine,University of Thessaly
Bloch, Karin Markenroth (författare)
Philips Healthcare Netherlands
Heiberg, Einar (författare)
Lund University,Lunds universitet,Klinisk fysiologi, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Clinical Physiology (Lund),Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
Carlsson, Marcus (författare)
Lund University,Lunds universitet,Klinisk fysiologi, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Clinical Physiology (Lund),Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
Arheden, Håkan (författare)
Lund University,Lunds universitet,Klinisk fysiologi, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Clinical Physiology (Lund),Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
visa färre...
 (creator_code:org_t)
2015-06-18
2015
Engelska.
Ingår i: BMC Medical Imaging. - : Springer Science and Business Media LLC. - 1471-2342. ; 15:1
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Background: Respiratory gating is often used in 4D-flow acquisition to reduce motion artifacts. However, gating increases scan time. The aim of this study was to investigate if respiratory gating can be excluded from 4D flow acquisitions without affecting quantitative intracardiac parameters. Methods: Eight volunteers underwent CMR at 1.5 T with a 5-channel coil (5ch). Imaging included 2D flow measurements and whole-heart 4D flow with and without respiratory gating (Resp(+), Resp(-)). Stroke volume (SV), particle-trace volumes, kinetic energy, and vortex-ring volume were obtained from 4D flow-data. These parameters were compared between 5ch Resp(+) and 5ch Resp(-). In addition, 20 patients with heart failure were scanned using a 32-channel coil (32ch), and particle-trace volumes were compared to planimetric SV. Paired comparisons were performed using Wilcoxon's test and correlation analysis using Pearson r. Agreement was assessed as bias ± SD. Results: Stroke volume from 4D flow was lower compared to 2D flow both with and without respiratory gating (5ch Resp(+) 88 ± 18 vs 97 ± 24.0, p = 0.001; 5ch Resp(-) 86 ± 16 vs 97.1 ± 22.7, p < 0.01). There was a good correlation between Resp(+) and Resp(-) for particle-trace derived volumes (R2 = 0.82, 0.2 ± 9.4 ml), mean kinetic energy (R2 = 0.86, 0.07 ± 0.21 mJ), peak kinetic energy (R2 = 0.88, 0.14 ± 0.77 mJ), and vortex-ring volume (R2 = 0.70, -2.5 ± 9.4 ml). Furthermore, good correlation was found between particle-trace volume and planimetric SV in patients for 32ch Resp(-) (R2 = 0.62, -4.2 ± 17.6 ml) and in healthy volunteers for 5ch Resp(+) (R2 = 0.89, -11 ± 7 ml), and 5ch Resp(-) (R2 = 0.93, -7.5 ± 5.4 ml), Average scan duration for Resp(-) was shorter compared to Resp(+) (27 ± 9 min vs 61 ± 19 min, p < 0.05). Conclusions: Whole-heart 4D flow can be acquired with preserved quantitative results without respiratory gating, facilitating clinical use.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Kardiologi (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Cardiac and Cardiovascular Systems (hsv//eng)

Nyckelord

4D flow
Cardiac MRI
Kinetic energy
Particle trace
Respiratory gating
Vortex ring size

Publikations- och innehållstyp

art (ämneskategori)
ref (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy