A new small-angle X-ray scattering set-up on the crystallography beamline I711 at MAX-lab.

Knaapila, Matti (författare)
Lund University,Lunds universitet,MAX IV-laboratoriet,MAX IV Laboratory
Svensson, Christer (författare)
Lund University,Lunds universitet,MAX IV-laboratoriet,MAX IV Laboratory
Barauskas, Justas (författare)
Lund University,Lunds universitet,Fysikalisk kemi,Enheten för fysikalisk och teoretisk kemi,Kemiska institutionen,Institutioner vid LTH,Lunds Tekniska Högskola,Physical Chemistry,Physical and theoretical chemistry,Department of Chemistry,Departments at LTH,Faculty of Engineering, LTH
visa fler...
Zackrisson Oskolkova, Malin (författare)
Lund University,Lunds universitet,Fysikalisk kemi,Enheten för fysikalisk och teoretisk kemi,Kemiska institutionen,Institutioner vid LTH,Lunds Tekniska Högskola,Physical Chemistry,Physical and theoretical chemistry,Department of Chemistry,Departments at LTH,Faculty of Engineering, LTH
Nielsen, S S (författare)
Toft, K N (författare)
Vestergaard, B (författare)
Arleth, L (författare)
Olsson, Ulf (författare)
Lund University,Lunds universitet,Fysikalisk kemi,Enheten för fysikalisk och teoretisk kemi,Kemiska institutionen,Institutioner vid LTH,Lunds Tekniska Högskola,Physical Chemistry,Physical and theoretical chemistry,Department of Chemistry,Departments at LTH,Faculty of Engineering, LTH
Pedersen, J S (författare)
Cerenius, Yngve (författare)
Lund University,Lunds universitet,MAX IV-laboratoriet,MAX IV Laboratory
visa färre...
 (creator_code:org_t)
2009
2009
Engelska.
Ingår i: Journal of Synchrotron Radiation. - 1600-5775. ; 16:Pt 4, s. 498-504
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • A small-angle X-ray scattering (SAXS) set-up has recently been developed at beamline I711 at the MAX II storage ring in Lund (Sweden). An overview of the required modifications is presented here together with a number of application examples. The accessible q range in a SAXS experiment is 0.009-0.3 A(-1) for the standard set-up but depends on the sample-to-detector distance, detector offset, beamstop size and wavelength. The SAXS camera has been designed to have a low background and has three collinear slit sets for collimating the incident beam. The standard beam size is about 0.37 mm x 0.37 mm (full width at half-maximum) at the sample position, with a flux of 4 x 10(10) photons s(-1) and lambda = 1.1 A. The vacuum is of the order of 0.05 mbar in the unbroken beam path from the first slits until the exit window in front of the detector. A large sample chamber with a number of lead-throughs allows different sample environments to be mounted. This station is used for measurements on weakly scattering proteins in solutions and also for colloids, polymers and other nanoscale structures. A special application supported by the beamline is the effort to establish a micro-fluidic sample environment for structural analysis of samples that are only available in limited quantities. Overall, this work demonstrates how a cost-effective SAXS station can be constructed on a multipurpose beamline.

Ämnesord

NATURVETENSKAP  -- Fysik -- Acceleratorfysik och instrumentering (hsv//swe)
NATURAL SCIENCES  -- Physical Sciences -- Accelerator Physics and Instrumentation (hsv//eng)

Publikations- och innehållstyp

art (ämneskategori)
ref (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy