Sökning: id:"swepub:oai:lup.lub.lu.se:32fba28d-7081-476a-82f9-8f4e89569b8f" >
Rapid diagnostic te...
Rapid diagnostic testing for SARS-CoV-2 : Validation and comparison of three point-of-care antibody tests
-
- Strand, Rasmus (författare)
- Lund University
-
- Thelaus, Louise (författare)
- Lund University,Lunds universitet,Translationell Sepsisforskning,Forskargrupper vid Lunds universitet,Translational Sepsis research,Lund University Research Groups
-
- Fernström, Nils (författare)
- Lund University
-
visa fler...
-
- Sunnerhagen, Torgny (författare)
- Lund University,Lunds universitet,Translationell infektionsmedicin,Forskargrupper vid Lunds universitet,Translational infection medicine,Lund University Research Groups
-
- Lindroth, Ylva (författare)
- Lund University,Lunds universitet,Avdelningen för medicinsk mikrobiologi,Institutionen för laboratoriemedicin,Medicinska fakulteten,Division of Medical Microbiology,Department of Laboratory Medicine,Faculty of Medicine
-
- Linder, Adam (författare)
- Lund University,Lunds universitet,Translationell Sepsisforskning,Forskargrupper vid Lunds universitet,Translational Sepsis research,Lund University Research Groups
-
- Rasmussen, Magnus (författare)
- Lund University,Lunds universitet,Translationell infektionsmedicin,Forskargrupper vid Lunds universitet,Translational infection medicine,Lund University Research Groups
-
visa färre...
-
(creator_code:org_t)
- 2021-05-03
- 2021
- Engelska 5 s.
-
Ingår i: Journal of Medical Virology. - : Wiley. - 0146-6615 .- 1096-9071. ; 93:7, s. 4592-4596
- Relaterad länk:
-
https://portal.resea... (primary) (free)
-
visa fler...
-
http://dx.doi.org/10... (free)
-
https://onlinelibrar...
-
https://lup.lub.lu.s...
-
https://doi.org/10.1...
-
visa färre...
Abstract
Ämnesord
Stäng
- With the emergence of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), a need for diagnostic tests has surfaced. Point-of-care (POC) antibody tests can detect immunoglobulin (Ig) G and M against SARS-CoV-2 in serum, plasma, or whole blood and give results within 15 min. Validation of the performance of such tests is needed if they are to be used in clinical practice. In this study, we evaluated three POC antibody tests. Convalescent serum samples from 47 reverse transcription-polymerase chain reaction (RT-PCR) verified patients with coronavirus disease 2019 (COVID-19) collected at least 28 days post RT-PCR diagnosis as well as 50 negative pre-COVID-19 controls were tested. The three tests (denoted the J-, N-, and Z-tests) displayed the sensitivities of 87%, 96%, and 85%, respectively, for the detection of IgG. All tests had the same specificity for IgG (98%). The tests did not differ significantly for the detection of IgG. The sensitivities for IgM were lower (15%, 67%, and 70%) and the specificities were 90%, 98%, and 90%, respectively. The positive and negative predictive values were similar among the tests. Our results indicate that these POC antibody tests might be accurate enough to use in routine clinical practice.
Ämnesord
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Infektionsmedicin (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Infectious Medicine (hsv//eng)
Nyckelord
- COVID-19
- point-of-care test
- SARS-CoV-2
- sensitivity
- specificity
- validation