Sökning: id:"swepub:oai:lup.lub.lu.se:34fb05cc-dcf7-4db2-83db-73d4b2097db4" >
Methodological cons...
Methodological considerations on the determination of the APC response in plasma
-
- Andersson, N. E. (författare)
- Chromogenix AB
-
- Lindberg, K. (författare)
- Chromogenix AB
-
- Johansson, K. (författare)
- Chromogenix AB
-
visa fler...
-
- Rosen, S. (författare)
- Chromogenix AB
-
- Mahlke, K. (författare)
- Chromogenix AB
-
- Rabe, K F (författare)
- Chromogenix AB
-
- Zoller, B. (författare)
- Lund University,Lunds universitet,Allmänmedicin och klinisk epidemiologi,Forskargrupper vid Lunds universitet,Family Medicine and Clinical Epidemiology,Lund University Research Groups,Skåne University Hospital
-
- Dahlback, B. (författare)
- Lund University,Lunds universitet,Klinisk kemi, Malmö,Forskargrupper vid Lunds universitet,Clinical Chemistry, Malmö,Lund University Research Groups,Skåne University Hospital
-
visa färre...
-
(creator_code:org_t)
- 2009-05-27
- 1995
- Engelska 3 s.
-
Ingår i: Infusionstherapie und Transfusionsmedizin. - : S. Karger AG. - 1019-8466. ; 22:SUPPL. 1, s. 80-82
- Relaterad länk:
-
http://dx.doi.org/10...
-
visa fler...
-
https://lup.lub.lu.s...
-
https://doi.org/10.1...
-
visa färre...
Abstract
Ämnesord
Stäng
- The performance of COATEST® APC® Resistance on different coagulation instruments has been further evaluated through analysis of plasma from 100 blood donors on KC 10, ST 4, ACL 300R, Electra 900 and Thrombolyzer. The electromechanical instruments KC-10 and ST 4 showed a lower response to APC than the turbidimetric or photometric instruments with median APC ratios of 2.7 and 2.8 for the two former versus 3.1, 3.4 and 3.5, respectively, for the other three, which is in agreement with earlier initial findings. Similarly, the cut-off value varied between 2.1 and 2.6 for these instruments. The correlation of APC ratios between instruments was strong with r values ranging between 0.71 and 0.93 and, furthermore, none of the six plasmas with the lowest APC ratios on the Thrombolyzer ranked higher than 8 on any of the other instruments. Analysis of control plasmas with six consecutive kit batches on ACL and ST4 resulted in APC ratio ranges of 3.3-3.7 and 2.7-3.0 for a normal control and 1.8-2.0 and 1.9-2.0 for an abnormal control on ACL and ST4, respectively, illustrating a high reproducibility between batches. Repeated freezing and thawing of samples is disrecommended since this often resulted in increased APC ratios. In contrast, in spite of up to 40% decrease in FVIII activity upon storage of 10 different plasma samples for 5 h, the effect on the APC ratio was only minor as was also the effect of addition of 1.0 IU/ml of FVIII. In neither case was any sample misclassified. Altogether, the results support the applicability of this kit for measuring the response of plasma to APC.
Ämnesord
- MEDICIN OCH HÄLSOVETENSKAP -- Medicinska och farmaceutiska grundvetenskaper -- Läkemedelskemi (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Basic Medicine -- Medicinal Chemistry (hsv//eng)
Nyckelord
- activated protein C (APC)
- APC response in plasma
Publikations- och innehållstyp
- art (ämneskategori)
- ref (ämneskategori)
Hitta via bibliotek
Till lärosätets databas