Repeatability of T1-quantification in dGEMRIC for three different acquisition techniques: two-dimensional inversion recovery, three-dimensional look locker, and three-dimensional variable flip angle.

Siversson, Carl (författare)
Lund University,Lunds universitet,Enheten för led- och mjukvävnadsforskning,Forskargrupper vid Lunds universitet,Medicinsk strålningsfysik, Malmö,Joint and Soft Tissue Unit,Lund University Research Groups,Medical Radiation Physics, Malmö
Tiderius, Carl Johan (författare)
Lund University,Lunds universitet,Enheten för led- och mjukvävnadsforskning,Forskargrupper vid Lunds universitet,Joint and Soft Tissue Unit,Lund University Research Groups
Neuman, Paul (författare)
Lund University,Lunds universitet,Enheten för led- och mjukvävnadsforskning,Forskargrupper vid Lunds universitet,Joint and Soft Tissue Unit,Lund University Research Groups
visa fler...
Dahlberg, Leif (författare)
Lund University,Lunds universitet,Enheten för led- och mjukvävnadsforskning,Forskargrupper vid Lunds universitet,Joint and Soft Tissue Unit,Lund University Research Groups
Svensson, Jonas (författare)
Lund University,Lunds universitet,Enheten för led- och mjukvävnadsforskning,Forskargrupper vid Lunds universitet,Medicinsk strålningsfysik, Malmö,Joint and Soft Tissue Unit,Lund University Research Groups,Medical Radiation Physics, Malmö
visa färre...
 (creator_code:org_t)
Wiley, 2010
2010
Engelska.
Ingår i: Journal of Magnetic Resonance Imaging. - : Wiley. - 1522-2586 .- 1053-1807. ; 31:5, s. 1203-1209
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • PURPOSE: To evaluate the repeatability of the dGEMRIC (delayed gadolinium enhanced MRI of cartilage) method in osteoarthritis-prone knee joints for three different T1 quantification techniques: two-dimensional inversion recovery (2D-IR), three-dimensional Look-Locker (3D-LL), and three-dimensional variable flip angle (3D-VFA). MATERIALS AND METHODS: Nine subjects were examined twice, with a 2-week interval, using all three measurement techniques. Four regions of interest were defined in the central medial and lateral femoral cartilage. The repeatability was evaluated for each measurement technique. For the 3D techniques, the variation between different slices was also evaluated. RESULTS: Repeatability expressed by root-mean-square coefficient of variation (CV(RMS)) showed similar results for 2D-IR and 3D-LL (5.4-8.4%). For 3D-VFA CV(RMS) was higher (9.3-15.2%). Intraclass correlation coefficient showed both 2D-IR and 3D-LL reliability to be moderate, while 3D-VFA reliability was low. Inter-slice CV(RMS) and ICC was of the same magnitude as the repeatability. No clear differences could be interpreted between the condyles. CONCLUSION: Both 2D-IR and 3D-LL perform well in generating repeatable dGEMRIC results, while 3D-VFA results are somewhat inferior. Furthermore, repeatability results in this study are similar to previously published results for healthy subjects. Finally, the positioning of the analyzed images is crucial to generate reliable repeatability results.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Publikations- och innehållstyp

art (ämneskategori)
ref (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy