SwePub
Sök i LIBRIS databas

  Utökad sökning

id:"swepub:oai:lup.lub.lu.se:6a6af119-8365-4e61-8b63-318fc75ebb71"
 

Sökning: id:"swepub:oai:lup.lub.lu.se:6a6af119-8365-4e61-8b63-318fc75ebb71" > European quadricent...

European quadricentric evaluation of a breast MR biopsy and localization device: technical improvements based on phase-I evaluation

Prat, X (författare)
Sittek, H (författare)
Grosse, A (författare)
visa fler...
Bååth, Lars (författare)
Lund University,Lunds universitet,Diagnostisk radiologi, Malmö,Forskargrupper vid Lunds universitet,Radiology Diagnostics, Malmö,Lund University Research Groups
Perlet, C (författare)
Alberich, T (författare)
Lamarque, JM (författare)
Andersson, Ingvar (författare)
Lund University,Lunds universitet,Diagnostisk radiologi, Malmö,Forskargrupper vid Lunds universitet,Radiology Diagnostics, Malmö,Lund University Research Groups
Reiser, M (författare)
Taourel, P (författare)
Fischer, H (författare)
Heywang-Kobrunner, SH (författare)
visa färre...
 (creator_code:org_t)
2002-03-23
2002
Engelska.
Ingår i: European Radiology. - : Springer Science and Business Media LLC. - 0938-7994 .- 1432-1084. ; 12:7, s. 1720-1727
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Our purpose was to report about technical success, problems and solutions, as experienced in a first multicentre study on MR-guided localisation or vacuum biopsy of breast lesions. The study was carried out at four European sites using a dedicated prototype breast biopsy device. Experiences with 49 scheduled localisation procedures and 188 vacuum biopsies are reported. Apart from 35 dropped indications, one localisation procedure and 9 vacuum biopsies were not possible (3 times space problems due to obesity, 2 times too strong compression, 3 times impaired access froth medially, 2 times impaired access due to a metal bar). Problems due to too strong compression were recognised by repeat MR without compression. During the procedure problems leading to an uncertain result occurred in eight vacuum biopsies, two related to the procedure: one limited access, and one strong post-biopsy enhancement. Improvements after phase-I study concerned removal of the metal bar, development of an improved medial access, of a profile imitating the biopsy gun, optimisation of compression plates and improved software support. The partners agreed that the improvements answered all important technical problems.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Nyckelord

breast biopsy
breast MR

Publikations- och innehållstyp

art (ämneskategori)
ref (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy