New device, ‘old’ allergens. Allergic contact dermatitis caused by the Dexcom G7 glucose sensor

Mowitz, Martin (författare)
Lund University,Lunds universitet,Yrkes- och miljödermatologi,Forskargrupper vid Lunds universitet,Dermatologi,Occupational and Environmental Dermatology,Lund University Research Groups,Dermatology,Skåne University Hospital
Hosseini, Sarah (författare)
Skåne University Hospital,Lund University,Ryhov County Hospital, Jönköping
Siemund, Ingrid (författare)
Lund University,Lunds universitet,Yrkes- och miljödermatologi,Forskargrupper vid Lunds universitet,Occupational and Environmental Dermatology,Lund University Research Groups,Skåne University Hospital
visa fler...
Ulriksdotter, Josefin (författare)
Lund University,Lunds universitet,Yrkes- och miljödermatologi,Forskargrupper vid Lunds universitet,Occupational and Environmental Dermatology,Lund University Research Groups,Skåne University Hospital
Svedman, Cecilia (författare)
Lund University,Lunds universitet,Yrkes- och miljödermatologi,Forskargrupper vid Lunds universitet,Occupational and Environmental Dermatology,Lund University Research Groups,Skåne University Hospital
visa färre...
 (creator_code:org_t)
Engelska.
Ingår i: Contact Dermatitis. - 0105-1873.
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Background: Allergic contact dermatitis (ACD) has been reported as an adverse effect from the use of several glucose sensors and insulin pumps from different manufacturers. Isobornyl acrylate (IBOA) has been identified as a major culprit sensitizer, but also other acrylates and (modified) colophonium have been reported as causes of ACD. Objectives: To report the two first cases diagnosed with ACD caused by the Dexcom G7 (DG7) glucose sensor. Patients and Methods: Two children with suspected ACD from DG7 were patch tested with our medical device series with an addition of selected test preparations including two variants of modified colophonium – methyl hydrogenated rosinate (MHR) and glyceryl hydrogenated rosinate (GHR). Both patients were also tested with acetone extracts made from different parts of the DG7 sensor. The extracts were analysed by gas chromatography–mass spectrometry (GC–MS). Results: Both patients tested positive to IBOA, hydroabietyl alcohol and GHR. In addition, patient 1 had a positive reaction to MHR and patient 2 had a positive reaction to colophonium. The GC–MS analyses showed the presence of IBOA and colophonium-related substances in the DG7 extracts. Conclusions: Both patients were diagnosed with contact allergy to well-known medical device-related sensitizers. The presence of IBOA and (modified) colophonium in a newly introduced (on the Swedish market in 2023) glucose sensor is remarkable and indicates an inadequate toxicological assessment of the materials used in the sensor.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Dermatologi och venereologi (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Dermatology and Venereal Diseases (hsv//eng)

Nyckelord

allergic contact dermatitis
case report
colophonium
Dexcom G7
glucose sensor
glyceryl hydrogenated rosinate
hydroabietyl alcohol
isobornyl acrylate
medical device
methyl hydrogenated rosinate
rosin

Publikations- och innehållstyp

art (ämneskategori)
ref (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy