Evaluation of the relationship between capillary and venous plasma glucose concentrations obtained by the HemoCue Glucose 201+ system during an oral glucose tolerance test

Ignell, Claes (författare)
Lund University,Lunds universitet,Genomik, diabetes och endokrinologi,Forskargrupper vid Lunds universitet,Genomics, Diabetes and Endocrinology,Lund University Research Groups
Berntorp, Kerstin (författare)
Lund University,Lunds universitet,Genomik, diabetes och endokrinologi,Forskargrupper vid Lunds universitet,Genomics, Diabetes and Endocrinology,Lund University Research Groups
 (creator_code:org_t)
2011-10-03
2011
Engelska.
Ingår i: Scandinavian Journal of Clinical & Laboratory Investigation. - : Informa UK Limited. - 1502-7686 .- 0036-5513. ; 71:8, s. 670-675
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Abstract In 55 women with previous gestational diabetes mellitus, simultaneous capillary and venous plasma glucose concentrations were measured at 0, 30 and 120 min during a 75 g oral glucose tolerance test (OGTT). The aims of the study were to examine the relationship between capillary and venous glucose measurements, and to establish equations for the conversion of capillary and venous glucose concentrations using the HemoCue Glucose 201+ system. Additionally, the correlation between the capillary and venous glucose concentrations with the diagnostic cut-off limits proposed by the World Health Organization (WHO) in 1999 was evaluated. Capillary glucose concentrations were consistently higher than venous glucose concentrations at all time points of the OGTT (p < 0.001), and the correlations between the measurements were statistically highly significant (p < 0.001). The differences between the samples were greatest in the non-fasting state as revealed by the 95% prediction intervals (mmol/L) in Bland-Altman plots; ? 0.54 at 0 min, ? 2.01 at 30 min, and ? 1.35 at 120 min. Equivalence values for capillary plasma glucose concentrations derived from this study tended to be higher than those proposed by the WHO as diagnostic cut-off limits. Stratifying subjects by glucose tolerance status according to the WHO criteria revealed disagreements related to glucose values close to the diagnostic cut-off points. The study findings highlight the uncertainty associated with derived equivalence values. However, capillary plasma glucose measurements could be suitable for diagnostic purposes in epidemiological studies and when translating results on a group basis.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Endokrinologi och diabetes (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Endocrinology and Diabetes (hsv//eng)

Nyckelord

correlation
conversion algorithm
comparison method
Agreement
Bland-Altman plot

Publikations- och innehållstyp

art (ämneskategori)
ref (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Hitta mer i SwePub

Av författaren/redakt...
Ignell, Claes
Berntorp, Kersti ...
Om ämnet
MEDICIN OCH HÄLSOVETENSKAP
MEDICIN OCH HÄLS ...
och Klinisk medicin
och Endokrinologi oc ...
Artiklar i publikationen
Scandinavian Jou ...
Av lärosätet
Lunds universitet

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy