APM Task Group 329 : Reference dose specification for dose calculations: Dose-to-water or dose-to-muscle?

Kry, Stephen F. (författare)
University of Texas
Feygelman, Vladimir (författare)
H. Lee Moffitt Cancer Center & Research Institute
Balter, Peter (författare)
University of Texas
visa fler...
Knöös, Tommy (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Radiotherapy Physics,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Lund,Section V,Department of Clinical Sciences, Lund,Faculty of Medicine,Lund University Research Groups,Skåne University Hospital
Charlie Ma, C. M. (författare)
Fox Chase Cancer Center
Snyder, Michael (författare)
Wayne State University
Tonner, Brian (författare)
Ackerman Cancer Center
Vassiliev, Oleg N. (författare)
University of Texas
visa färre...
University of Texas H Lee Moffitt Cancer Center & Research Institute (creator_code:org_t)
2020-01-28
2020
Engelska 13 s.
Ingår i: Medical Physics. - : Wiley. - 0094-2405 .- 2473-4209. ; 47:3, s. 52-64
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Linac calibration is done in water, but patients are comprised primarily of soft tissue. Conceptually, and specified in NRG/RTOG trials, dose should be reported as dose-to-muscle to describe the dose to the patient. Historically, the dose-to-water of the linac calibration was often converted to dose-to-muscle for patient calculations through manual application of a 0.99 dose-to-water to dose-to-muscle correction factor, applied during the linac clinical reference calibration. However, many current treatment planning system (TPS) dose calculation algorithms approximately provide dose-to-muscle (tissue), making application of a manual scaling unnecessary. There is little guidance on when application of a scaling factor is appropriate, resulting in highly inconsistent application of this scaling by the community. In this report we provide guidance on the steps necessary to go from the linac absorbed dose-to-water calibration to dose-to-muscle in patient, for various commercial TPS algorithms. If the TPS does not account for the difference between dose-to-water and dose-to-muscle, then TPS reference dose scaling is warranted. We have tabulated the major vendors' TPS in terms of whether they approximate dose-to-muscle or calculate dose-to-water and recommend the correction factor required to report dose-to-muscle directly from the TPS algorithm. Physicists should use this report to determine the applicable correction required for specifying the reference dose in their TPS to achieve this goal and should remain attentive to possible changes to their dose calculation algorithm in the future.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Nyckelord

calibration
medium
muscle
reference
water

Publikations- och innehållstyp

art (ämneskategori)
ref (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy