Arterio-venous blood gas Δvalues for validation of umbilical cord blood samples at birth are not only biased by sample mix ups but also affected by clinical factors
-
- Mokarami, Parisa (författare)
- Lund University,Lunds universitet,Urogynekologi och reproduktionsfarmakologi,Forskargrupper vid Lunds universitet,Urogynaecology and Reproductive Pharmacology,Lund University Research Groups,Skåne University Hospital
-
- Wiberg, Nana (författare)
- Lund University,Lunds universitet,Obstetrisk, gynekologisk och prenatal ultraljudsdiagnostik,Forskargrupper vid Lunds universitet,Obstetric, Gynaecological and Prenatal Ultrasound Research,Lund University Research Groups,Skåne University Hospital
-
- Källén, Karin (författare)
- Lund University,Lunds universitet,Tornbladinstitutet,Forskargrupper vid Lunds universitet,Tornblad Institute,Lund University Research Groups
-
visa fler...
-
- Olofsson, Per (författare)
- Lund University,Lunds universitet,Urogynekologi och reproduktionsfarmakologi,Forskargrupper vid Lunds universitet,Urogynaecology and Reproductive Pharmacology,Lund University Research Groups,Skåne University Hospital
-
visa färre...
-
(creator_code:org_t)
- 2018-10-30
- 2019
- Engelska.
-
Ingår i: Acta Obstetricia et Gynecologica Scandinavica. - : Wiley. - 0001-6349. ; 98:2, s. 167-175
- Relaterad länk:
-
http://dx.doi.org/10...
-
visa fler...
-
https://obgyn.online...
-
https://lup.lub.lu.s...
-
https://doi.org/10.1...
-
visa färre...
Abstract
Ämnesord
Stäng
- Introduction: Traditional validation of umbilical cord blood samples with positive veno-arterial ΔpH and arterio-venous ΔpCO2 values confirms the source of samples, whereas negative Δvalues represent mix-up of samples. To investigate whether this is true, the distributions of V-A ΔpO2 and A-V Δlactate were also explored and related to clinical characteristics. In addition, different cord blood sampling techniques were evaluated. Material and methods: Register study with cord blood acid-base and clinical data from 27 233 newborns. Clinical characteristics were related to positive, zero and negative Δvalues. Blood samplings from unclamped and double-clamped cords were compared. A two-sided P < 0.05 was considered significant. Results: ΔpH and ΔpCO2 values distributed into positive, around zero, and negative sub-populations, with significant differences in pH and clinical characteristics between sub-populations. No such sub-populations were distinguished for ΔpO2 and Δlactate. The 2.5th and 5th ΔpH percentiles were 0.013 and 0.022, respectively, and for ΔpCO2 0.30 and 0.53 kPa. Applying 5th percentile criteria resulted in 3.5% of “approved” cases showing a ΔpO2 ≤ 0. Puncture and sampling of the unclamped cord resulted in significantly better sample quality. Conclusions: Unphysiological negative ΔpO2 values occurred despite correct validation with traditional criteria. Δlactate cannot be used for validation because both positive and negative values are physiological. Positive/around zero/negative ΔpH and ΔpCO2 sub-populations were associated with significant differences in pH and clinical characteristics, indicating that defective sampling and sample handling are not the sole explanations for negative Δvalues. Prompt puncture and sampling of the unclamped cord resulted in best sample quality.
Ämnesord
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Reproduktionsmedicin och gynekologi (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Obstetrics, Gynaecology and Reproductive Medicine (hsv//eng)
Nyckelord
- blood gases
- delayed sampling
- lactate
- pCO
- pH
- pO
- pregnancy
- umbilical cord
- validation