Calculation and measurements of absorbed dose in total body irradiation
-
- Svahn-Tapper, Gudrun (författare)
- Lund University,Lunds universitet,Medicinsk strålningsfysik, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Medical Radiation Physics, Lund,Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
-
- Nilsson, Per (författare)
- Lund University,Lunds universitet,Medicinsk strålningsfysik, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Medical Radiation Physics, Lund,Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
-
- Jönsson, Christel (författare)
- Lund University,Lunds universitet,Diagnostisk radiologi, Malmö,Forskargrupper vid Lunds universitet,Radiology Diagnostics, Malmö,Lund University Research Groups
-
visa fler...
-
- Alvegård, Thor (författare)
- Lund University,Lunds universitet,Medicinsk onkologi,Sektion I,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Medical oncology,Section I,Department of Clinical Sciences, Lund,Faculty of Medicine
-
visa färre...
-
(creator_code:org_t)
- 2009-07-08
- 1990
- Engelska.
-
Ingår i: Acta Oncologica. - : Informa UK Limited. - 1651-226X .- 0284-186X. ; 29:5, s. 627-633
- Relaterad länk:
-
http://dx.doi.org/10...
-
visa fler...
-
https://lup.lub.lu.s...
-
https://doi.org/10.3...
-
visa färre...
Abstract
Ämnesord
Stäng
- A method which is simple, reliable, and rapid to use in clinical routine for basic dose calculation in total body irradiation (TBI) has been tested with 8 MV x-rays. The dosimetry follows, as far as possible, national and international recommendations for conventional radiotherapy. The dose rate at different locations and depths is calculated from the absorbed dose rate at dose maximum for a phantom size of 30 x 30 x 30 cm in the TBI field (Dc), an inverse square law factor (SAD2/SPD2), the tissue-maximum ratio (TMR), an equivalent phantom and patient size correction factor (A), a factor for lack of back-scattering material (B), an off-axis output correction factor (O), and a factor that corrects for off-axis variations in effective photon beam energy and for oblique beam penetration of the patient (R). The collimator opening is constant for all patient sizes. It is shown that TMR, A, B and R can be measured in conventional geometry in ordinary phantoms but at an extended distance, while Dc, O and SAD2/SPD2 must be measured in TBI geometry. Tests in Humanoid phantoms showed an agreement in measured and planned AP/2 doses of 2% or better. If the calculation method is used for lower photon energies or in other TBI geometries it may be necessary to correct for the elliptical shape of the patient and for back-scattered radiation from the walls or floor.
Ämnesord
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Cancer och onkologi (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Cancer and Oncology (hsv//eng)
Nyckelord
- leukaemia
- 8 MV x-rays
- radiotherapy
- Total body irradiation
- dosimetry
- bone marrow transplantation
Publikations- och innehållstyp
- art (ämneskategori)
- ref (ämneskategori)