Sökning: id:"swepub:oai:lup.lub.lu.se:cea9fd61-b347-481a-b714-9ee63f6d5d03" >
Towards real-time m...
Towards real-time magnetomotive ultrasound imaging
-
- Evertsson, Maria (författare)
- Lund University,Lunds universitet,Medicinskt ultraljud,Forskargrupper vid Lunds universitet,Medical ultrasound,Lund University Research Groups
-
- Ramalli, Alessandro (författare)
- University of Florence
-
- Pavan, Theo Z. (författare)
- University of São Paulo
-
visa fler...
-
- Cabrelli, Luciana C. (författare)
- University of São Paulo
-
- Andersson, Roger (författare)
- Skåne University Hospital
-
- Cinthio, Magnus (författare)
- Lund University,Lunds universitet,Avdelningen för Biomedicinsk teknik,Institutionen för biomedicinsk teknik,Institutioner vid LTH,Lunds Tekniska Högskola,Department of Biomedical Engineering,Departments at LTH,Faculty of Engineering, LTH
-
- Tortoli, Piero (författare)
- University of Florence
-
- Jansson, Tomas (författare)
- Lund University,Lunds universitet,Medicinskt ultraljud,Forskargrupper vid Lunds universitet,Medical ultrasound,Lund University Research Groups
-
visa färre...
-
(creator_code:org_t)
- 2017
- 2017
- Engelska.
-
Ingår i: 2017 IEEE International Ultrasonics Symposium, IUS 2017. - 9781538633830
- Relaterad länk:
-
http://dx.doi.org/10...
-
visa fler...
-
https://lup.lub.lu.s...
-
https://doi.org/10.1...
-
visa färre...
Abstract
Ämnesord
Stäng
- Enabling detection of nanoparticles with ultrasound can open new application avenues for the ultrasound technique. Magnetomotive ultrasound (MMUS) is a technique under development which indirectly visualizes magnetic nanoparticles. In MMUS, an external time-varying magnetic field acts to displace the nanoparticles, and thus their closest surrounding. This induced displacement is subsequently detected and the nanoparticle location may then be revealed. The MMUS technique has shown to be promising in both phantom and animal studies but limited efforts have been made on optimizing the technique for clinical applications in the sense of providing real-time bedside imaging. In this work, the previously proposed MMUS algorithm is automated and implemented online on the ULA-OP scanner. To evaluate the online implementation, a phantom made of styrene-ethylene/butylene-styrene and mineral oil with a 2 % magnetic ferrite particle inclusion was used. MMUS displacement was calculated in the entire image area, 192×230 pixels, and in a sub-region of 130×90 pixels, covering the inclusion. It was found that the automated online implementation computes one full MMUS image in 2.8 seconds and the sub-region in 1.17 seconds, which should be compared to 1-2 minutes in post processing mode. An immediate on-screen change in the magnetomotive displacement could be observed as the applied magnetic field was altered.
Ämnesord
- TEKNIK OCH TEKNOLOGIER -- Medicinteknik -- Medicinsk bildbehandling (hsv//swe)
- ENGINEERING AND TECHNOLOGY -- Medical Engineering -- Medical Image Processing (hsv//eng)
Nyckelord
- Contrast agents
- Magnetomotive ultrasound imaging
- ULA-OP
Publikations- och innehållstyp
- kon (ämneskategori)
- ref (ämneskategori)
Hitta via bibliotek
Till lärosätets databas