Variable velocity encoding in a three-dimensional, three-directional phase contrast sequence: Evaluation in phantom and volunteers.

Nilsson, Anders (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Medical Radiation Physics, Lund,Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
Markenroth Bloch, Karin (författare)
Lund University,Lunds universitet,Klinisk fysiologi, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Clinical Physiology (Lund),Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
Carlsson, Marcus (författare)
Lund University,Lunds universitet,Klinisk fysiologi, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Clinical Physiology (Lund),Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
visa fler...
Heiberg, Einar (författare)
Lund University,Lunds universitet,Klinisk fysiologi, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Clinical Physiology (Lund),Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
Ståhlberg, Freddy (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Medical Radiation Physics, Lund,Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
visa färre...
 (creator_code:org_t)
2012-10-12
2012
Engelska.
Ingår i: Journal of Magnetic Resonance Imaging. - : Wiley. - 1522-2586 .- 1053-1807.
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • PURPOSE: To evaluate accuracy and noise properties of a novel time-resolved, three-dimensional, three-directional phase contrast sequence with variable velocity encoding (denoted 4D-vPC) on a 3 Tesla MR system, and to investigate potential benefits and limitations of variable velocity encoding with respect to depicting blood flow patterns. MATERIALS AND METHODS: A 4D PC-MRI sequence was modified to allow variable velocity encoding (VENC) over the cardiac cycle in all three velocity directions independently. 4D-PC sequences with constant and variable VENC were compared in a rotating phantom with respect to measured velocities and noise levels. Additionally, comparison of flow patterns in the ascending aorta was performed in six healthy volunteers. RESULTS: Phantom measurements showed a linear relationship between velocity noise and velocity encoding. 4D-vPC MRI presented lower noise levels than 4D-PC both in phantom and in volunteer measurements, in agreement with theory. Volunteer comparisons revealed more consistent and detailed flow patterns in early diastole for the variable VENC sequences. CONCLUSION: Variable velocity encoding offers reduced noise levels compared with sequences with constant velocity encoding by optimizing the velocity-to-noise ratio (VNR) to the hemodynamic properties of the imaged area. Increased VNR ratios could be beneficial for blood flow visualizations of pathology in the cardiac cycle. J. Magn. Reson. Imaging 2012. © 2012 Wiley Periodicals, Inc.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Publikations- och innehållstyp

art (ämneskategori)
ref (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy