Factors affecting the trapping performance of xenon holdup-filters in nuclear medicine applications

Bolmsjö, M. S. (författare)
Lund University
Persson, B. R.R. (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Medical Radiation Physics, Lund,Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
 (creator_code:org_t)
1998-06-04
1982
Engelska.
Ingår i: Medical Physics. - : Wiley. - 0094-2405. ; 9:1, s. 96-105
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Activated charcoal traps are used in nuclear medicine departments to capture exhaled radioactive xenon gas. In the present study, the trapping performance of charcoal was investigated for different physical qualities in experimental and clinical situations. Various factors affecting the trapping capacity are identified, such as the charcoal mass, moisture and carbonâ€�dioxide concentrations in the sweep gas, gasâ€�flow rate through the trap, and temperature. Improper drying of the exhaled air prior to passage through the charcoal bed appears to be a critical factor, leading to a much earlier breakthrough of xenon. Commercial xenon traps intended for ambient temperature operation were found to be useful only for a limited number of patients due to an early breakthrough point. By refrigerating the trap, however, a much larger capacity was achieved. If correctly managed, xenon traps constitute a useful alternative for isolating radioactive xenon wastes in clinical and experimental applications.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Nyckelord

ACTIVATED CARBON
ADSORPTION
CHARCOAL
FEASIBILITY STUDIES
FILTERS
PERFORMANCE
RADIOACTIVE WASTE DISPOSAL
REFRIGERATION
SORPTIVE PROPERTIES
TRAPS
XENON 133

Publikations- och innehållstyp

art (ämneskategori)
ref (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Hitta mer i SwePub

Av författaren/redakt...
Bolmsjö, M. S.
Persson, B. R.R.
Om ämnet
MEDICIN OCH HÄLSOVETENSKAP
MEDICIN OCH HÄLS ...
och Klinisk medicin
och Radiologi och bi ...
Artiklar i publikationen
Medical Physics
Av lärosätet
Lunds universitet

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy