Restoring Natural Forearm Rotation in Transradial Osseointegrated Amputees

Boni, Irene (författare)
Integrum AB
Millenaar, Jason (författare)
Integrum AB
Controzzi, Marco (författare)
Scuola Superiore Sant'Anna di studi universitari e di perfezionamento (SSSUP),Sant'Anna School of Advanced Studies (SSSUP)
visa fler...
Ortiz Catalan, Max Jair, 1982 (författare)
Integrum AB,Chalmers tekniska högskola,Chalmers University of Technology
visa färre...
 (creator_code:org_t)
2018
2018
Engelska.
Ingår i: IEEE Transactions on Neural Systems and Rehabilitation Engineering. - 1558-0210 .- 1534-4320. ; 26:12, s. 2333-2341
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Osseointegrated transradial prostheses have the potential to preserve the natural range of wrist rotation, which improves the performance of activities of daily living and reduces compensatory movements that potentially lead to secondary health problems over time. This is possible by enabling the radius and the ulna bone to move with respect to each other, restoring the functionality of the original distal-radioulnar joint. In this paper, we report on psychophysics tests performed on an osseointegrated transradial amputee with the aim to understand the extent of mobility of the implants that is required to preserve the natural forearm rotation. Based on these experiments, we designed and developed an attachment device between the implants and the hand prosthesis that serves as an artificial distal radio-ulnar joint. This device was fitted on an osseointegrated transradial amputee and its functionality assessed by means of the Southampton Hand Assessment Procedure (SHAP) and the Minnesota Manual Dexterity test (MMDT). We found that the axial rotation of the implants is required to preserve forearm rotation, to distribute loads equally over the two implants (60% radius - 40% ulna), and to enable loading of the implants without unpleasant feelings for the patient. Higher function was recorded when our attachment device enabled forearm rotation: SHAP from 61 to 71, MMDT from 258s to 231s. Natural forearm rotation can be successfully restored in transradial amputees by using osseointegration and our novel mechanical attachment to the hand prosthesis.

Ämnesord

TEKNIK OCH TEKNOLOGIER  -- Medicinteknik -- Annan medicinteknik (hsv//swe)
ENGINEERING AND TECHNOLOGY  -- Medical Engineering -- Other Medical Engineering (hsv//eng)
MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Odontologi (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Dentistry (hsv//eng)
MEDICIN OCH HÄLSOVETENSKAP  -- Medicinsk bioteknologi -- Biomedicinsk laboratorievetenskap/teknologi (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Medical Biotechnology -- Biomedical Laboratory Science/Technology (hsv//eng)

Nyckelord

transradial amputation
Artificial limbs
prosthesis fitting
osseointegration
forearm rotation

Publikations- och innehållstyp

art (ämneskategori)
ref (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Hitta mer i SwePub

Av författaren/redakt...
Boni, Irene
Millenaar, Jason
Controzzi, Marco
Ortiz Catalan, M ...
Om ämnet
TEKNIK OCH TEKNOLOGIER
TEKNIK OCH TEKNO ...
och Medicinteknik
och Annan medicintek ...
MEDICIN OCH HÄLSOVETENSKAP
MEDICIN OCH HÄLS ...
och Klinisk medicin
och Odontologi
MEDICIN OCH HÄLSOVETENSKAP
MEDICIN OCH HÄLS ...
och Medicinsk biotek ...
och Biomedicinsk lab ...
Artiklar i publikationen
IEEE Transaction ...
Av lärosätet
Chalmers tekniska högskola

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy