Quality assurance guidelines for superficial hyperthermia clinical trials: II. Technical requirements for heating devices

Dobsicek Trefna, Hana, 1979 (författare)
Chalmers tekniska högskola,Chalmers University of Technology
Crezee, J. (författare)
Universiteit Van Amsterdam,University of Amsterdam
Schmidt, M. (författare)
Universitatsklinik Erlangen und Medizinische Fakultat
visa fler...
Marder, D. (författare)
Lamprecht, U. (författare)
Eberhard Karls Universität Tübingen,Eberhard Karls University of Tübingen
Ehmann, M. (författare)
Nadobny, J. (författare)
Charité Universitätsmedizin Berlin,Charité University Medicine Berlin
Hartmann, J. (författare)
Universitatsklinik Erlangen und Medizinische Fakultat
Lomax, N. (författare)
Abdel-Rahman, S. (författare)
Ludwig-Maximilians-Universität München,Ludwig Maximilian University of Munich
Curto, S. (författare)
Erasmus Universiteit Rotterdam,Erasmus University Rotterdam
Bakker, A. (författare)
Universiteit Van Amsterdam,University of Amsterdam
Hurwitz, M. D. (författare)
Diederich, C. J. (författare)
University of California
Stauffer, P. R. (författare)
Van Rhoon, G. C. (författare)
Erasmus Universiteit Rotterdam,Erasmus University Rotterdam
visa färre...
 (creator_code:org_t)
2017-03-01
2017
Engelska.
Ingår i: Strahlentherapie und Onkologie. - : Springer Science and Business Media LLC. - 1439-099X .- 0179-7158. ; 193:5, s. 351-366
  • Forskningsöversikt (refereegranskat)
Abstract Ämnesord
Stäng  
  • Quality assurance (QA) guidelines are essential to provide uniform execution of clinical trials with uniform quality hyperthermia treatments. This document outlines the requirements for appropriate QA of all current superficial heating equipment including electromagnetic (radiative and capacitive), ultrasound, and infrared heating techniques. Detailed instructions are provided how to characterize and document the performance of these hyperthermia applicators in order to apply reproducible hyperthermia treatments of uniform high quality. Earlier documents used specific absorption rate (SAR) to define and characterize applicator performance. In these QA guidelines, temperature rise is the leading parameter for characterization of applicator performance. The intention of this approach is that characterization can be achieved with affordable equipment and easy-to-implement procedures. These characteristics are essential to establish for each individual applicator the specific maximum size and depth of tumors that can be heated adequately. The guidelines in this document are supplemented with a second set of guidelines focusing on the clinical application. Both sets of guidelines were developed by the European Society for Hyperthermic Oncology (ESHO) Technical Committee with participation of senior Society of Thermal Medicine (STM) members and members of the Atzelsberg Circle.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine (hsv//eng)
TEKNIK OCH TEKNOLOGIER  -- Medicinteknik (hsv//swe)
ENGINEERING AND TECHNOLOGY  -- Medical Engineering (hsv//eng)

Nyckelord

Applicator
Quality assurance
Hyperthermia
Phantoms
superficial
Water bolus
Heating criteria

Publikations- och innehållstyp

for (ämneskategori)
ref (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy