SwePub
Sök i LIBRIS databas

  Utökad sökning

id:"swepub:oai:research.chalmers.se:20de857e-926f-4d82-854f-8d9543e145f1"
 

Sökning: id:"swepub:oai:research.chalmers.se:20de857e-926f-4d82-854f-8d9543e145f1" > Towards an electrow...

Towards an electrowetting-based digital microfluidic platform for magnetic immunoassays

Schaller, Vincent, 1979 (författare)
Chalmers tekniska högskola,Chalmers University of Technology
Sanz-Velasco, Anke, 1971 (författare)
Chalmers tekniska högskola,Chalmers University of Technology
Kalaboukhov, Alexei, 1975 (författare)
Chalmers tekniska högskola,Chalmers University of Technology
visa fler...
Schneiderman, Justin, 1979 (författare)
Chalmers tekniska högskola,Chalmers University of Technology
Öisjöen, Fredrik, 1982 (författare)
Chalmers tekniska högskola,Chalmers University of Technology
Jesorka, Aldo, 1967 (författare)
Chalmers tekniska högskola,Chalmers University of Technology
Prieto Astalan, Andrea (författare)
Krozer, Anatol, 1949 (författare)
Rusu, Cristina (författare)
Enoksson, Peter, 1957 (författare)
Chalmers tekniska högskola,Chalmers University of Technology
Winkler, Dag, 1957 (författare)
Chalmers tekniska högskola,Chalmers University of Technology
visa färre...
 (creator_code:org_t)
Royal Society of Chemistry (RSC), 2009
2009
Engelska.
Ingår i: Lab on a Chip - Miniaturisation for Chemistry and Biology. - : Royal Society of Chemistry (RSC). - 1473-0189 .- 1473-0197. ; 9:23, s. 3433-3436
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • We demonstrate ElectroWetting-On-Dielectric (EWOD) transport and SQUID gradiometer detection of magnetic nanoparticles (MNPs) suspended in a 2 mu l de-ionized water droplet. This proof-of-concept methodology constitutes the first development step towards a highly sensitive magnetic immunoassay platform with SQUID readout and droplet-based sample handling. Magnetic AC-susceptibility measurements were performed on MNPs with a hydrodynamic diameter of 100 nm using a high-Tc dc Superconducting Quantum Interference Device (SQUID) gradiometer as detector. We observed that the signal amplitude per unit volume is 2.5 times higher for a 2 ml sample droplet compared to a 30 ml sample volume.

Ämnesord

TEKNIK OCH TEKNOLOGIER  -- Medicinteknik -- Medicinsk laboratorie- och mätteknik (hsv//swe)
ENGINEERING AND TECHNOLOGY  -- Medical Engineering -- Medical Laboratory and Measurements Technologies (hsv//eng)
TEKNIK OCH TEKNOLOGIER  -- Materialteknik -- Annan materialteknik (hsv//swe)
ENGINEERING AND TECHNOLOGY  -- Materials Engineering -- Other Materials Engineering (hsv//eng)
TEKNIK OCH TEKNOLOGIER  -- Maskinteknik -- Strömningsmekanik och akustik (hsv//swe)
ENGINEERING AND TECHNOLOGY  -- Mechanical Engineering -- Fluid Mechanics and Acoustics (hsv//eng)
NATURVETENSKAP  -- Fysik -- Den kondenserade materiens fysik (hsv//swe)
NATURAL SCIENCES  -- Physical Sciences -- Condensed Matter Physics (hsv//eng)

Publikations- och innehållstyp

art (ämneskategori)
ref (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy